2.1.1. Apply a static preset
Note: Attribute 105 can only be written using implicit messaging
•
As implicit messaging is not supported by the tool we will write attribute 19 via explicit
messaging.
•
Set the data field to 0xe8030000 to apply a preset to 1000 (still using mm resolution)
•
Observe that the display will show 1000mm
•
Write 0x01 to attribute 106 to reset this preset
Note: Attribute 106 monitors 0-1 transitions. Thus, you might need to write 0x00 in case you have
previously written 0x01.
© SICK AG • Subject to change without notice • 8016993/ZVZ7/2018-03-22
Betriebsanleitung Entfernungs-Messgerät DL100 Pro – EtherNet/IP