Herunterladen Inhalt Inhalt Diese Seite drucken

Vapormed Mighty MEDIC Gebrauchsanweisung Seite 196

Vorschau ausblenden Andere Handbücher für Mighty MEDIC:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
10. E L VA P O R I Z A D O R M I G H T Y M E D I C
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MIGHT Y MEDIC Vaporizer is intended for use in the electromagnetic en-
vironment specified below. The customer or the user of the MIGHT Y MEDIC
Vaporizer should assure that it is used in such an environment.
Immunity
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
test level
3 V
rms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
Compliance level Electromagnetic environ-
3 V
rms
3 V/m
196
ment – guidance
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the MIGHTY
MEDIC Vaporizer,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance:
d = 1,2 √P
d = 1,2 √P
für 80 MHz to 800 MHz
d = 2,3 √P
for 800 MHz to 2,5 GH