10. E L VA P O R I Z A D O R M I G H T Y M E D I C
10.5. Información con Respecto a la Compatibilidad electromagnética
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The MIGHT Y MEDIC Vaporizer is intended for use in the electromagnetic en-
vironment specified below. The customer or the user of the MIGHT Y MEDIC
Vaporizer should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment - guidance
Group 1
The MIGHT Y MEDIC Vaporizer uses
RF energy only for its internal func-
tion. Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby electronic
equipment.
Class B
The MIGHT Y MEDIC Vaporizer is suit-
able for use in all establishments, in-
cluding domestic establishments and
Class A
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Complies