Herunterladen Inhalt Inhalt Diese Seite drucken

Setting The Custom Program - MTR+ Biolito Bedienungsanleitung

Inkontinenztherapie-stimulator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1. Switch on Biolito by pressing the „On/Off" button.
2. Choose the program PC06 with the „P" button.
3. Press the „S" button for at least three seconds. Biolito changes to the programming mode
and the time that indicates program duration, flashes to show that this function can be altered.
In the following please use the left „+" button to store your choice and to change to the
next step. The right „+/-" buttons are respectively used in order to alter the chosen and
flashing parameter.
4. Select with the right „+/-"buttons the desired time for the phase.
5. By pressing the left „+"button you will be able to store your choice and change to the next
step. Mode now flashes on the LCD and you are able to select the mode that you require by
pressing the right „+/-"buttons.
If you have chosen „CONT" or „BURST" you can set the frequency (Hz) and pulse width
(µS).
If you have chosen „MOD" you can set the following parameters:
- upper limit of frequency (Hz HI)
- low limit of frequency (Hz LO)
- upper limit of pulse width (µS HI)
- low limit of pulse width ( µS LO)
- low limit of intensity (AdJ) - the display in the right corner below shows values from 0.5 to
1.0. This means that the intensity of current can increase and decrease within the modulation
time between 50% (=0.5) and 100% (=1.0) of the originally set mA-value. The upper limit of
the intensity (100%) is always the mA value adjusted by the user originally. The intensity in a
modulated program, can never increase above the 100% of the originally set intensity.
- modulation time (TM - possible choice: 2 sec. up to 60 sec)
Programming of Dense-Disperse-Stimulation according to Prof. Han (CONT + BURST) is not
possible.
Biolito Bedienungsanleitung / Operators Manual
Setting the
Custom Program (06)
38