Myolito Bedienungsanleitung / Operators Manual
Setting the
Custom Program (12)
1. Switch on Myolito with the "On/Off" button.
2. Choose the program PC12 with the "P" button.
3. Press the "S" button for at least three seconds. Myolito changes to the
programming mode and the time that indicates program duration, flashes to
show that this function can be altered.
In the following please use the left "+" button always then, if you want
to store your choice and to change to the next step. The right
"+/-" buttons are respectively used in order to alter the chosen and
flashing parameter.
4. Select with the right "+/-"buttons the desired time for the phase.
5. By pressing the left "+"button you will be able to store your choice and
change to the next step. Mode now flashes on the LCD and you are able to
select the mode that you require by pressing the right "+/-"buttons.
If you have chosen "CONT" or "BURST" you can set the frequency (Hz)
and pulse width (µS).
If you have chosen "MOD" you can set the following parameters:
- upper limit of frequency (Hz HI)
- low limit of frequency (Hz LO)
- upper limit of pulse width (µS HI)
- low limit of pulse width ( µS LO)
- low limit of intensity (AdJ) - the display in the right corner below shows
values from 0.5 to 1.0. This means that the intensity of current can increase
and decrease within the modulation time between 50% (=0.5) and 100% (=1.0)
of the originally set mA-value. The upper limit of the intensity (100%) is
always the mA value adjusted by the user originally. The intensity in a
modulated program, can never increase above the 100% of the originally set
intensity.
- modulation time (TM - possible choice: 2 sec. up to 60 sec)
32