Myolito Bedienungsanleitung / Operators Manual
LOCK-button and Statistics
The Myolito offers the possibility to define one program and exclude the
others. This allows the doctor or therapist to ensure that the patient uses only
the program that has been selected for him.
For this purpose the LOCK-button in the battery-compartment has to be
pressed with a small pin once (the metal-pins of the electrode-cables are well
suited). The lock symbol appears on the display and the program cannot be
changed by pressing the "P"-button anymore. Pressing the LOCK-button a
second time allows reading the following statistics for the treatment in the
display:
Total Treatment Time
Number of all sessions
Lock-Symbol
Average used intensity
per channel
If the program was not locked, it is necessary to press the LOCK button twice
to read the memory,. The statistics are recorded in general, regardless whether
the program was defined or not.
To empty the memory, the "S"-button must be pressed after reading for three
seconds continuously. This action will also unlock the program and all the
individual settings in program PC12 are reset to their default values.
36