Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Specification - Facom E.2000-50 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 5
Technical Specifi cation
Weight:
E.2000-50 / E.2000-350 : 2,6 Kg
Dimensions:
E.2000-50 / E.2000-350 : 145,5 mm deep x 150 mm wide x 85 mm high.
Accuracy:
+/-1%, +/-1 digit of the fi gure read.
Display:
4 Digit, 7 segment LED. With update rate of twice per second (2Hz).
Temp Range - Operating: +5°C ➜ +40°C
-20°C ➜ +70°C
Temp Range - Storage:
Auto Reset Hold Time:
2 seconds
Electromagnetic Compatibility (EMC) Directive: In conformance with EN 61326:1997.
INSTALLATION
WARNINGS
• ENSURE THE MOUNTING SURFACE IS CAPABLE OF SUPPORTING THE INSTRUMENT WHEN: 50 N.m (E.2000-50), 350 N.m
(E.2000-350), 1000 N.m (E.2000-1000) TORQUE IS APPLIED.
• ONLY USE THE POWER SUPPLY PROVIDED.
• DO NOT APPLY TORQUE ABOVE: 50 N.m (E.2000-50), 350 N.m (E.2000-350), 1000 N.m (E.2000-1000).
1. Identify a suitable surface and position to mount the instrument.
2. Using the template on this guide, mark the position of the 4 mounting holes.
3. Drill (drill and tap) the 4 mounting holes suitable for 10.0 mm (3/8 inch) fasteners (M10; Grade 8.8 minimum; cap head type are recommended or the
equivalent 3/8 inch).
4. Fix instrument in position using fasteners at a torque of 35 to 40 N.m (25 to 30 ft.lb).
5. Connect power supply (supplied) to the instrument. All display segments will light and buzzer will sound briefl y. The instrument is ready for use.
OPERATION
MODE Button
1. Press MODE button briefl y to display current mode of operation.
2. Press and hold MODE button to change mode of operation. Release button at required mode to accept.
3. Modes of operation.
• Track Mode ('trAC' is displayed). Display follows torque applied
• Click Mode ('CLIC' is displayed). Display shows fi rst peak of torque applied. Serial data output of peak torque is automatically transmitted.
Display automatically resets after 3 seconds.
For use with 'click' type torque wrenches.
• Dial Mode ('dIAL' is displayed).
Display shows peak of torque applied. Press RESET to clear the display.
For use with 'dial' and 'electronic' type torque wrenches.
UNIT Button
Press UNIT to change units of measurement.
RESET Button
Press RESET to reset display in Dial mode. Serial data output of peak torque is also transmitted.
Press RESET to transmit serial data in Track mode.
LIMIT Button
1. Press LIMIT to enter target torque setting. Display fl ashes between 'SEt' and the target torque setting.
• Press MODE to increase target setting. Hold to increase speed of change.
• Press UNIT to decrease target setting. Hold to increase speed of change.
2. Press LIMIT to enter ± percentage tolerance setting from 1 to 10%. Enter '0' for no limits. '%' and the ± percentage tolerance setting is displayed constantly.
• Press MODE to increase ± percentage tolerance setting.
• Press UNIT to decrease ± percentage tolerance setting.
3. Press LIMIT to fi nish. Display shows 'Fin' for 1 second.
Using the Instrument
1. Select mode of operation.
2. Place wrench in the instrument and operate wrench in desired direction. Remove wrench and zero the display (if required) by pressing the RESET and
LIMIT buttons together. 'SEt0' is displayed for 1 second.
3. Place wrench in the instrument and operate wrench in desired direction.
Serial Port: Connect lead (supplied) between TruCheck Plus serial port and computer/ printer.
Communication settings: 9600 Baud, 8 data bits, 1 stop bit, no parity.
NU-E.2000_0609.indd 4
NU-E.2000_0609.indd 4
E.2000-1000 : 4,6 kg
E.2000-1000 : 170,5 mm x 175 mm x 110 mm
A
EN
4
18/09/2009 16:39:07
18/09/2009 16:39:07

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

834540E.2000-350834574E.2000-10002000

Inhaltsverzeichnis