Parameter list
Tab. 23
No.
Designation
3
Actual rotor
frequency
5
Actual pump
current
7
Actual motor
temperature
11
Actual
controller
temperature
12
Operating
mode for
Start/Stop
18
Nominal pump
frequency
24
Setpoint
frequency
125
Current bearing
temperature
127
Current bearing
temperature
184
Controller
operating
hours counting
303
Pump status
word
Special TV 301-P Navigator Controller User Manual / 87-901-002-01 (B)
Range
Unit
Default
0...1300
Hz
-
0...60
0.1A
-
-
°C
-
0...150
°C
-
0...1
-
0
750...1200
Hz
*
150...963
Hz
*
0...150
°C
-
0...150
°C
-
0...19
0.01
-
years
h
-
-
-
Technical Information
Format
r/w
Description
U16
r
The max driving frequency
U16
r
Measured pump current
U16
r
Returns the bearing
temperature
U16
r
Measured internal controller
temperature
U16
r/w
P12 = 0 (default): via remote
(J6)
P12 = 1 via serial interface
U16
r
Highest permissible motor
current
U16
r/w
Adjustable between 150 and
963 Hz
U16
r
Measured bearing temperature
(Identical to P127)
U16
r
Measured bearing temperature
(Identical to P125)
U32
r
Totals the operating ours of the
converter when the pump's
drive is active
U16
r
Meaning of the bits:
Bit 0 = Normal operation
Bit 1 = Ready for switch on
Bit 2 = Ramp
Bit 3 = Braking
In case of error P303 has the
value of 0
17
Profibus
171/180