Oxford
®
Elara Ceiling Lift
Lifting Tape Inspection and Renewal
The lifting tape must be inspected for wear, damage, fraying, cuts, discoloration and loose
or damaged threads by the user/caregiver BEFORE EVERY USE. If any deterioration is
detected DO NOT USE THE LIFT and contact an authorised Joerns Healthcare service
provider to arrange replacement.
Lift Tape Replacement
As part of the bi-annual inspection schedule, the lifting tape should be inspected for wear,
damage, fraying, cuts, discoloration and loose or damaged threads and replaced as necessary.
Regardless of its condition, Joerns Healthcare recommend the lifting tape is replaced every 5
years or 5000 lift cycles, whichever comes first.
Packing & Delivery
In the unlikely event of future breakdown, it is recommended that packaging is retained. This will
help minimise the possibility of transit damage if the need to return goods arises.
Equipment Identification
The equipment identification (serial number, model number and manufacturing date) and
specification (safe working load, electrical rating/electrical details in accordance with IEC 60601-
1:2006 + A12:2014, manufacturer's details) are detailed on the label affixed to the unit's moulded
cabinet.
6
WARNING
English