Oxford
®
Elara Ceiling Lift
Displays, Settings and Menus

1. Standby Mode

In standby mode, the following information is displayed:
NOTE: If the lift is not placed on charge following 'LOW BATTERY' alert, the LCD display panel will
indicate 'BATTERY DEPLETED' and no further lifts will be possible.
BATTERY CHARGE STATUS SYMBOL
The battery charge status symbol indicates the remaining charge when the lift unit is in standby mode.
The indicator either shows 1/3 rd , 2/3 rd or fully charged.
Refer to the 'Charge / Status Indicator' section for more details.
When the battery level falls to a critical level (reaches the point at which only one full lift is
achievable) the charge / status indicator will illuminate permanent red (not flashing). This will
be accompanied by the alert 'LOW BATTERY' on the LCD display screen, Ensure the lift unit is
returned to charge immediately.
2. Sleep Mode
To protect the batteries from unnecessary discharge, the ceiling lift enters sleep mode when no
control input is detected after a period of 1 minute. The display will awake when any of the operating
buttons are depressed.