Use of the battery charger

2.2.1 Battery charger

Primary voltage
Secondary voltage (charging voltage)
Secondary current (charging current)
Safeguards
Efficiency
Ambient temperature
Length of mains lead
Ambient storage temperature
Relative storage humidity
Conformity

2.2.2 Charging the batteries

Use only the battery charger supplied and no other charging equipment.
We recommend that you should regularly recharge the built-in control box batteries in order
to ensure that the hoist remains usable and to extend the lifespan of the batteries. The
control box will give the warning sound to tell you when the battery charge is too low.
SETTING UP THE CHARGER
When setting up the charger, ensure that it is well-ventilated on all sides. A minimum of
100 mm space should be left free around the unit for this purpose. If the charger is
insufficiently ventilated so that the unit heats up, the charging rate will fall which will extend
the charging time. If the charger overheats (> +50°C), it will stop charging.
The charger should only be used from a wall socket with a voltage of 100 to 240V at
50/60Hz and in well-ventilated, dry areas.
FIRST USE
First, insert the recharge cable into the appropriate connector on the control box. Then
insert the mains adapter into the socket. The loading time is about 4 hours. The charger will
stop the recharging process automatically when the device is fully loaded.
RECHARGING
WARNING: Risk of injury – Do not use the lift when the charger is

connected in the control box.

Eagle 620

100 - 240VAC / 50/60 Hz
27.6VDC (+/- 2%)
max. 500 mA
Protected against reverse polarisation, electrical
surges and extreme temperature
min. 80% (when fully loaded)
0 to +40° C
2.0 m
-15° to +50° C
max. 95% (non-condensing)
tested in accordance with IEC 60601-1
When the batteries are recharging there will burn a yellow
light on the upper LED.
After the recharging process is completed, you should always
remove the mains plug of the charger from the socket first
and then remove the charger from the control box.
Do not use the lift while it is being recharged!
If the batteries are not used for a significant period, they
discharge slowly by themselves
(deep discharge). It then becomes impossible to recharge
them with the battery charger supplied. You should therefore
recharge the batteries at least once a month even if they are
not being used.