The product 1 has a coupling, which allows for the attachment of various
saw blades 11. These saw blades 11 lock automatically when attached to
the product.
The saw blade 11 can be removed by activating a push button 13.
5.

Preparation

Aesculap assumes no liability if the following rules are not followed:
Do not use products from open or damaged sterile packaging.
Prior to use, inspect the product and its accessories for any visible
damage.
Use the products and their accessories only if they are in perfect tech-
nical condition.
6.

Working with the device

Risk of infection and contamination!
Product is delivered unsterilized!
WARNING
Sterilize the product before use pursuant to the
operating instructions.
Risk of injury and material damage due to acciden-
tal activation of the product!
Products which are not being actively used must
WARNING
be secured against accidental activation (posi-
tion OFF).
Risk of injury and material damage due to inappro-
priate use of tools!
WARNING
Always follow the safety advice and information
given in the instructions for use.
When coupling/uncoupling, handle tools with
cutting edges with care.
Damage to the product if dropped!
Use the products only if they are in perfect
technical condition, see Function check
WARNING
Risk of burns to skin and tissue through blunt tools
or if product has not been maintained properly!
Use tools only if they are in perfect condition.
WARNING
Replace blunt tools.
Maintain the product properly, see maintenance
guide.
6.1

System set-up

6.1.1

Connecting the accessories

Risk of injury due to unapproved configuration
using additional components!
DANGER
Ensure that the classification of all components
used is consistent with the classification of the
product (such as type BF or type CF).
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
All
configurations
must
IEC/DIN EN 60601-1. The person connecting the devices with each other
is responsible for the configuration and must ensure that the fundamental
standard IEC/DIN EN 60601-1 or relevant national standards.
Follow the instructions for use of individual accessories.
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
6.1.2

Inserting the rechargeable battery

The product does not function when using the
wrong battery!
CAUTION
Do not use battery GA666 (with red base) or
GA346.
Product 1 with the battery shaft 4 must be turned upwards and the
sterile funnel 6 (sterile) attached, see Fig. A.
Battery 7 (non-sterile) is inserted into the battery shaft 4 by a second
(non-sterile) person, see Fig. A.
Note
After the battery has been inserted, multiple signals will sound, which indi-
cate that the product is ready to use.
After insertion of the battery, have the Sterile funnel 6 (non-sterile)
removed by a second person.
Attach the cap 8 (sterile) in such a manner that it clicks into place with
both release catches 9.
Note
The sterility of the product is only guaranteed with a correctly placed cap.
fulfill