11. Technical data of the measuring accessories

-
Standard: EN 61010-031,
-
Maximum rated voltage to earth () and measuring category:
With push-on caps: 1000 V CAT III, 600 V CAT IV,
Without push-on caps: 1000 V CAT II,
-
Maximum rated current: 10 A,
-
Protective class II (  ), continuous double or reinforced insulation,
-
Contamination class: 2,
-
Length: 1.4 m, AWG 18,
-
Environmental conditions:
Maximum barometric elevation for making measurements: 2000 m,
Temperatures: 0 °C to + 50 °C, humidity 50 % to 80 %
-
Only use the measuring leads if in perfect and clean condition as well as
according to this manual, since the protection provided could otherwise be
impaired.
-
Throw the measuring lead out if the insulation is damaged or if there is a
break in the lead/ plug.
-
Do not touch the bare contact tips of the measuring lead. Only grab the area
appropriate for hands!
-
Insert the angled terminals in the testing or measuring device.
12. Environmental notice
At the end of the product's useful life, please dispose of it at
appropriate collection points provided in your country.