First remove the safety measuring leads from the measurement circuit.
-
Remove the safety measuring leads from the BENNING MM 12.
-
Turn the rotating switch J to "OFF".
-
Remove the protective rubber holster O from the BENNING MM 12.
-
Put the BENNING MM 12 face down and unscrew the lower screw of the battery compart-
ment cover.
-
Lift the battery compartment cover off the bottom part.
-
Lift one end of the defective fuse from the fuse holder.
-
Push the defective fuse out of the fuse holder completely.
-
Replace the defective fuse with another of the same rated power, same triggering charac-
teristics and same dimensions.
-
Push the new fuse into the centre of the holder.
-
Lock the battery compartment cover into place on the bottom part and tighten the screw.
-
Place the BENNING MM 12 into the rubber protection frame O .
See figure 13:
9.5 Calibration
Benning guarantees compliance with the technical and accuracy specifications stated in the
operating manual for the first 12 months after the delivery date.
To maintain the specified accuracy of the measurement results, the instrument must be
recalibrated at regular intervals by our factory service. We recommend a recalibration interval of
one year. Send the multimeter to the following address:
Benning Elektrotechnik & Elektronik GmbH & CO. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt
9.6 Spare parts
Fuse F 440 mA, 1000 V, 10 kA, D = 10 mm, L = 34.9 mm,
Fuse F 11 A, 1000 V, 20 kA, D = 10.3 mm, L = 38.1 mm,
10. How to use the protective rubber holster
-
The safety measuring leads can be stored by coiling them round the protective rubber
holster O and clipping the probe into the holster O so that they are sufficiently protected
(see fig. 14).
-
You can clip one lead onto the protective rubber holster O in such a way that the measuring
probe projects. This allows you to bring the measuring probe and the BENNING MM 12 up
to the measuring point together.
-
The support at the back of the holster O can be used to prop the BENNING MM 12 up in
a diagonal position (to make reading easier) or to suspend it (see fig. 15).
-
The protective rubber holster O has an eyelet for suspending the unit in a convenient posi-
tion.
See figure 14:
See figure 15:
11. Technical data of the measuring accessories
-
Standard: EN 61010-031,
-
Maximum rated voltage to earth () and measuring category:
With push-on caps: 1000 V CAT III, 600 V CAT IV,
Without push-on caps: 1000 V CAT II,
-
Maximum rated current: 10 A,
-
Protective class II (), continuous double or reinforced insulation,
-
Contamination class: 2,
-
Length: 1.4 m, AWG 18,
-
Environmental conditions:
Maximum barometric elevation for making measurements: 2000 m,
Temperatures: 0 °C to + 50 °C, humidity 50 % to 80 %
-
Only use the measuring leads if in perfect and clean condition as well as according to this
manual, since the protection provided could otherwise be impaired.
-
Throw the measuring lead out if the insulation is damaged or if there is a break in the lead/
plug.
-
Do not touch the bare contact tips of the measuring lead. Only grab the area appropriate
for hands!
-
Insert the angled terminals in the testing or measuring device.
12. Environmental notice
At the end of the product's useful life, please dispose of it at appropriate collection
points provided in your country.
12/ 2018
Fuse replacement
Winding up the safety measuring leads
Standing up the BENNING MM 12
BENNING MM 12
P.no. 10016655
P.no. 10016656
33