Ensure the isolating transformer is only connected to a power supply with a functioning
protective earth connection that complies with the provisions of the IEC 60364-7-710 2012-10
"Electrical installations in buildings, part 7-710: Requirements for special installations
or locations – Medical locations". If in doubt, contact a specialist in the electrical trade or an
authorised employee of the hospital's engineering team.
Personnel (hospital and service personnel) working directly or indirectly with a mobile
equipment carrier must be instructed!
Setting adjustments may only be carried out by qualified personnel.
Repairs may only be carried out by ITD GmbH.
Safe working with the equipment cart:
The device is only disconnected from the mains supply when the power plug is removed
from the socket.
Operation:
With every change of location it must be ensured that no one is injured or objects damaged!
Connections:
In the USA and Canada, use a hospital-grade connection cable when connecting isolating
transformers in the 115 V mode and in Japan use the optional Japanese connection cable.
When connecting to the multiple-socket strip, secure the plug using an optional available
plug pull-out lock.
Only devices that fulfil the requirements of IEC 60601-1 or are IEC-tested may be connected
to the sockets/connecting lines.
Additional medical equipment with connecting bolts for equipotential bonding are to be
connected with the green-yellow cable to the optional equipotential bonding connecting bolts!
Caution: The overall rating given on the nameplate must not be exceeded.
Please observe that no further multiple sockets must be connected to an existing
multiple socket.
Load capacity:
The total weight of the equipment and the accessories on the mobile equipment carrier may
not exceed the permitted payload weight (see load capacity sticker on the base frame).
The surface load imprinted on the system components may not be exceeded!
The load shown on the fittings (e.g. infusion stand, jointed arms) may not be exceeded!
Caution: Please note that, in accordance with the standard, the overall dimensi-
ons of the equipment cart, including all devices and systems, must be specified
on a label that is attached to the equipment cart. We would be pleased to provide
you with any support in creating this label.
Infection protection:
Hygiene regulations are to be observed when cleaning!
Give only cleaned and disinfected equipment and fittings to a service technician for
maintenance and repair work!
Environmental protection:
Dispose of all cleaning and disinfection agent residue in a manner not harmful to the
environment!
IMPORTANT INFORMATION
Operating manual for mobile equipment carriers 09/2020