Herunterladen Inhalt Inhalt Diese Seite drucken

Endress+Hauser Proline Prosonic Flow P 500 Installationsanleitung Seite 29

Austausch sensor, sensorhalterung, installations set
Vorschau ausblenden Andere Handbücher für Proline Prosonic Flow P 500:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

8.4.2
Installation for measurement via 2 traverses
Requirements
• The installation distance is known →  9
• Strapping bands are pre-assembled
Material
The following material is required for mounting:
• Two strapping bands incl. mounting bolts and centering plates where necessary (already
pre-assembled →  19, →  21)
• A mounting rail to position the strapping bands:
• Short rail up to DN 200 (8")
• Long rail up to DN 600 (24")
• No rail > DN 600 (24"), as distance measured by sensor distance between the mounting
bolts
• Two mounting rail holders
• Two sensor holders
• Coupling medium (coupling pad or coupling gel) for an acoustic connection between the
sensor and pipe
• Two sensors incl. connecting cables
• Open-ended wrench (13 mm)
• Screwdriver
Procedure:
1.
Position the strapping bands using the mounting rail [only DN50 to 600 (2 to 24"), for
larger nominal diameters, measure the distance between the center of the strap bolts
directly]: Fit the mounting rail with the bore identified by the letter (from the Result
sensor distance / measuring aid parameter) over the mounting bolt of strapping band
1 that is fixed in place. Position the adjustable strapping band 2 and fit the mounting
rail with the bore identified by the numerical value over the mounting bolt.
 17
Determining the distance according to the mounting rail (e.g. G22)
2.
Tighten strapping band 2 so that it cannot slip.
3.
Remove the mounting rail from the mounting bolt.
Endress+Hauser
G
22
EA01368D
29