EA01368D
9.
Position the two strap bolts on strapping band 1, arranging them at a half
circumference (180° arrangement, e.g. half past seven and half past one) or a quarter
circumference (90° arrangement, e.g. 10 o' c lock and 7 o' c lock) in relation to each other.
10. Tighten strapping band 1 so that it cannot slip.
11. Strapping band 2: proceed as for strapping band 1 (steps 4 to 8).
12. Only slightly tighten strapping band 2 for final mounting so that it can still be adjusted.
The distance/offset from the center of strapping band 2 to the center of strapping band
1 is indicated by the sensor distance of the device.
13. Align strapping band 2 so that it is perpendicular to the measuring pipe axis and
parallel to strapping band 1.
14. Position the two strap bolts on tensioning strap 2 on the measuring pipe so they are
parallel to one another and offset at the same height/clock position (e.g. 10 and 4
o' c lock) in relation to the two strap bolts on tensioning strap 1. A line drawn on the
measuring pipe wall that is parallel to the measuring pipe axis can be helpful here. Now
set the distance between the center of the strap bolts at the same level so that it exactly
matches the sensor distance. Alternatively, a wire length can be used → 26.
15. Tighten strapping band 2 so that it cannot slip.
WARNING
L
Risk of injury!
‣
To avoid sharp edges, trim the cut edges after shortening the strapping bands. Wear
suitable gloves and protective goggles.
7
Holder with strapping bands (large nominal diameters)
1
Strap bolt with guide*
2
Strapping band*
3
Tensioning screw
22
1
3
2
3
Endress+Hauser