Herunterladen Inhalt Inhalt Diese Seite drucken

Errors And Troubleshooting - Braun TempleSwipe BST200 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für TempleSwipe BST200:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 7
• When the user is learning how to use the thermometer for the first time until he/she has
familiarized himself/herself with the instrument and obtains consistent readings.
Memory mode
The Braun TempleSwipe™ Thermometer includes a memory feature to recall the last
temperature reading. The last temperature taken is stored in its memory and will be
automatically displayed for 2 seconds when it is turned on.
Temperature unit conversion and mute function
This thermometer has the ability to switch temperature units from Celsius to Fahrenheit as well the
ability to mute the device. When the device is in mute mode, all audible sounds will be disabled.
The mute icon will be displayed on the LCD at all times when the device is in mute mode.
1. While the device is off, press and hold the
until SET appears on the screen.
2. Release the measurement button
temperature scale and mute option are shown. The selected icons
will flash on the screen to confirm the new setting, then the
thermometer will turn off automatically.
Calibration
This product has been calibrated at the time of manufacture. If used according to the use
instructions, periodic calibration is not required. This device is not intended as a substitute for
regular check-ups by your doctor, please consult your doctor if you have any doubt about the
temperature reading.
Inquiry
In Europe, any serious incident (e.g. death, life-threatening injury, surgical intervention, ...)
that has occured in relation to this thermometer should be reported to Kaz Europe (see Helen
of Troy Consumer Service Numbers) and to the competent authority of your member state.
REF, LOT and Serial numbers of your product must be provided as they are essential to record
and follow up your inquiry or claim. The LOT and Serial numbers are located in the battery
compartment. The production date is given by the LOT number and can be deciphered as
explained below:
The first 3 numeric digits in the LOT number represent the day of the year of manufacture. The
next 2 numeric digits represent the last two numbers of the calendar year of manufacture and
the letter(s) at the end designates the production site. (E.g.: LOT No.: 12313tav this product
was made on the day 123, year 2013 at production site "tav".)
Guarantee
Please read all instructions before attempting to use this device. Please retain the receipt as
proof of and date of purchase. The receipt must be presented when making any claim within
the relevant guarantee period. Any claim under the guarantee will not be valid without a
proof of purchase. Your device is guaranteed for Two Years (2 Years) from date of purchase. This
guarantee covers defects in materials or workmanship that occur under normal use; defective
devices meeting these criteria will be replaced free of charge.
The guarantee DOES NOT cover defects or damage resulting from abuse or failure to follow
the user instructions. The guarantee becomes void if the device is opened, tampered with, or
6
button for 6 seconds
when the desired
used with non-Braun branded parts or accessories, or if repairs are undertaken by
unauthorized persons.
Accessories and consumables are excluded from any warranty.
For support requests, please visit www.BraunHealthcare.com/uk_en/ or find contact
information at the end of this owner's manual.
This Guarantee is applicable to Europe, Russia, Middle East and Africa only.
UK Only: This does not affect your consumer statutory rights.

Errors and troubleshooting

Error message
Situation
Error symbol: during taking
measurement
A "Err" sign appears when the
temperature measured is BELOW
34 ºC or 93.2 °F.
The mark "Err" appears if the
operating temperature is ABOVE
42.2 ºC or 108 °F.
When ambient temperature is too
high or too low to take an accurate
reading, device will provide an "Err"
message with flashing up/down
arrow icon. The error message will
appear after start up and memory
recall.
Low battery
When 20% of battery life is left, the
display FLASHES the low battery
warning symbol, however, the
device can still work until the
battery life has 0% left.
Dead battery indication
If the flashing battery icon is the
only symbol shown on the display,
the device cannot work.
Blank display
Thermometer does not have power.
All segments display or
Blank display
System error.
English
Solution
When you receive this symbol, press
the power button to reset the
thermometer before taking a new
measurement.
When you receive this symbol, press
the power button to reset the
thermometer before taking a new
measurement. If the ambient
temperature is not within the
allowed range of
15 ºC - 40 ºC (59 ºF - 104.0 ºF), you
will not be able to take a reading.
Replace batteries. See instructions
on page 11.
Replace batteries. See instructions
on page 11.
Please check if the batteries have
been loaded correctly. Also check
polarity (<+> and <–>) of batteries.
Replace batteries. See instructions
on page 11.
If error persists, call customer
service.
7

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis