0408-600-705 Rev-AB
Cleaning
WARNINGS:
• ALWAYS remove the power pack, any type power cord, earphone, and comfort
pads from the helmet before cleaning.
• DO NOT sterilize any component.
CAUTIONS:
• DO NOT use solvents, lubricants, or other chemicals, unless otherwise
specified.
• DO NOT allow liquids or moisture into the fan or electrical connections of the
helmet.
• DO NOT allow debris to enter the helmet fan.
NOTES:
• The cord may be the power cord or the dictation power cord.
• To clean the power pack, see the instructions for use supplied with the Flyte
charger.
To Clean the Surgical Helmet and Cord
1. Remove all the accessories from the helmet. Discard the comfort pads.
2. Wipe the helmet, cord, and power pack holster with a soft cloth dampened
with a nonabrasive, hospital disinfectant.
3. Dry the equipment with a lint-free towel or medical-grade compressed air.
4. Clean the headlight lens with lens tissue. See the instructions for use supplied
with the lens tissue.
5. See the Storage and Handling section.
Storage and Handling
CAUTION: ALWAYS store the equipment within the specified environmental
condition values throughout its useful life. See the Specifications section.
To ensure the longevity, performance and safety of this equipment, use of the
original packaging material is recommended when storing or transporting this
equipment.
Disposal/Recycle
WARNING: ALWAYS follow the current local recommendations and/or
regulations governing environmental protection and risks associated
with recycling or disposing of the equipment at the end of its useful
life.
To comply with European Community Waste Electrical and
Electronic Equipment (WEEE) Directive 2012/19/EU, this
device should be collected separately for recycling. Do
not dispose of as unsorted municipal waste. Contact local
distributor for disposal information. Ensure infected equipment
is decontaminated prior to recycling.
www.stryker.com
Troubleshooting
WARNING: DO NOT service this equipment, unless otherwise
specified. If you require service, contact your Stryker sales
representative or call Stryker customer service. Outside the US,
contact your nearest Stryker subsidiary.
PROBLEM
CAUSE
The face shield lens is
Not enough airflow is
fogging.
circulating throughout
the helmet.
The power pack requires
replacement.
The helmet is damaged.
The indicator lights fail
The power pack requires
to flash and/or the fan
replacement.
fails to turn on during
start up.
The power cord with
holster is damaged.
The helmet is damaged.
The helmet will not allow
The helmet is damaged.
airflow adjustment.
The fan fails to provide
The fan speed is not
enough or any airflow.
adjusted correctly.
The helmet is damaged.
The green microphone
The power pack may be
indicator flashes rapidly
dislodged or the fan may
for several seconds.
be experiencing physical
interference.
The steps within the To
The red power loss
indicator light flashes on
Don a Surgical Helmet
and off.
section were performed
out of sequence.
The power pack requires
replacement.
Sporadic electrical
Electrical noise is
interference is
present.
experienced.
NOTE: If any component must be discarded, see the Disposal/Recycle section.
EN
ACTION
Press the variable
control buttons to
increase the fan speed
if possible.
Replace the power pack
with a fully charged
power pack.
Replace the helmet.
Replace the power pack
with a fully charged
power pack.
Replace the power cord
with holster.
Replace the helmet.
Replace the helmet.
Press the variable
control buttons to
increase the fan speed.
Replace the helmet.
Make sure the power
pack is secure in the
holster. Remove any fan
obstruction.
See the To Don a
Surgical Helmet section
and perform the steps in
sequence.
Replace the power pack
with a fully charged
power pack to prevent
the loss of air circulation
in the helmet.
Turn off all the electrical
equipment not in use in
the room.
Relocate the electrical
equipment to maximize
the distance between
the equipment. Increase
spatial distance.
Plug equipment into
different outlets.
7