Herunterladen Inhalt Inhalt Diese Seite drucken

System Information; Error Log; Calibrate Touchscreen; Clean Centrifuge - Medtronic autoLog IQ Benutzerhandbuch

Autotransfusionssystem
Vorschau ausblenden Andere Handbücher für autoLog IQ:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 55
Size: 7 x 9 in (178 x 229 mm)
Table 2. Notifications, audible and visual alert signals (continued)
Notification
The SD card failed. See error log for
details. Contact Medtronic for service.
The vacuum system failed. See error log
for details. Contact Medtronic for service.
Storage available for only __ more case
records. Export and delete records
before they are overwritten.
The GUI communications failed. See
error log for details. Contact Medtronic for
service.
The keypad failed. See error log for
details. Contact Medtronic for service.
Figure 39. Notifications screen

19.5 System information

To view system information (software versions and runtime statistics) press System information from the Service and
maintenance screen.

19.6 Error log

To view device errors, press Error log from the Service and maintenance screen.

19.7 Calibrate touchscreen

Note: This function is not available during blood processing.
Note: If the touchscreen is inoperable, press and hold
on the screen after 10 s.
1. Press
to stop processing if necessary.
2. Wait until the device is fully stopped. The message Machine stopped. Press
3. Press Calibrate touchscreen from the Service and maintenance screen.
4. Follow the instructions on the screen.

19.8 Clean centrifuge

For instructions for cleaning the centrifuge, press Clean centrifuge from the Service and maintenance screen. Follow the
instructions on the screen. See Chapter 16 for full cleaning instructions.

19.9 Update software

Contact your Medtronic service representative to perform software updates.

19.10 Export service information

1. To export information for service, press Export service information from the Service and maintenance screen.
2. Insert a USB flash drive into a USB port (type A).
Caution: Use only a USB flash drive provided by Medtronic.
3. Press Download to USB.
4. The message Download is complete. appears. Press OK.
Note: If the download is unsuccessful, the message Download failed appears.
English
M967226A006 1B Medtronic Confidential
Description
The service log is not functional. Contact your
Medtronic service representative.
The vacuum system is not functional. Contact your
Medtronic service representative.
Device internal data storage contains 90 or more
case records. Delete records as described. See
Section 14.5.
There is a problem with the internal electronics.
Contact your Medtronic service representative.
The
or
control is not functioning properly.
Contact your Medtronic service representative.
for 10 s to calibrate the touchscreen. Follow the instructions that appear
50
Audible alert
Visual alert
signal
signal
None
Low
None
Low
None
Low
None
Low
None
Low
to enable self-start. appears.
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018
502431-132

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis