SEITE 8 - 148 x 210 mm - 3100021789/01 - SCHWARZ - 20-053 (kn)

Troubleshooting

Error
Cause
The charge level
The remaining battery charge
display on the hand-
is adequate only for typical 5
piece glows red
10-sec exposure cycles.
steadily.
The power level
The battery lacks sufficient
display of the hand-
charge.
piece flashes red.
The ongoing exposure
is interrupted (light off
signal is emitted) fol -
lowed by a 2-sec error
signal; the handpiece
switches to "sleep"
mode and resists
further activation.
The power level
Charging problem.
display of the hand-
The battery is defective or at
piece flashes red
the end of its useful life.
while the handpiece
is in the charger.
The handpiece has not
There is not enough charge
been used for a long
in the battery to turn on the
time and now it cannot
handpiece.
be turned on.
The light emission
The information about inter-
does not start when
mittent operation under the
the START button is
header "Technical Data -
pressed; an error
Handpiece" has not been
signal sounds for
observed.
2 sec.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
8
Solution
Place the handpiece in the
charger and re-charge the
battery.
Place the handpiece in the
charger and re-charge the
battery.
Replace the battery.
Place the handpiece in the
charger and re-charge the
battery.
Allow the handpiece to
cool for 3 minutes, then
start the next exposure
by pressing the START
button.
Error
Cause
During light emission
The information about inter-
in continuous mode,
mittent operation under the
an error signal sounds
header "Technical Data -
for 2 sec, the emission
Handpiece" has not been
is stopped, and the
observed.
handpiece changes to
The handpiece has become
sleep mode.
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
Allow the handpiece to
cool for 3 minutes, then
start the next exposure by
pressing the START button.
The LED showing the
The charge contact pins in
operating status in
the handpiece or the charger
the charger glows red
are wet.
steadily.
Dry the charge contact
pins. Be careful not to
bend the flexible charge
contact pins in the charger.
The LED showing the
Charger is defective.
operating status in the
Have the charger repaired.
charger flashes alter-
nately red and green.
The handpiece is not
in the charger.
The LED showing the
Malfunction during charging.
operating status in the
The battery is damaged.
charger flashes alter-
Replace the battery.
nately red and green.
The handpiece is in
the charger.
The LED showing the
Power outlet carries no
operating status in the
voltage.
charger does not come
Use a different power
on although the plug
outlet.
has been connected to
an electrical outlet.
Charger is defective.
Have the charger repaired.
The light intensity is
Clean the light guide and
too low.
the protecting glass in the
light guide mounting hole
(please refer to "Cleaning
the Light Guide").
Solution