Note!
• If a pause has been programmed, the
carriage will stop for the pause each
time the maximum stretching value is
attained.
• Enabling the stretching in supination
function automatically disables mo-
tor A. The only movement performed
is pronation/supination.
• It is not possible to combine a second
stretching function with the special
function "stretching in supination".
• During operation, the display will
show the symbol
left-hand corner.
LEVEL 6:
n EROM repeat
extension/pronation
The special "EROM repeat extension/
pronation" function allows a more ef-
ficient exercise in the last 10 ° before the
set maximum extension and pronation
values.
For this protocol, the CPM device starts
in the middle between the set maximum
extension/flexion and pronation/supi-
nation values. First of all the carriage
moves simultaneously to the maximum
flexion angle and to the maximum supi-
nation angle. Then it moves simultane-
ously to the maximum extension angle
and to the maximum pronation angle.
When the set extension/pronation value
has been reached, the carriage reverses
10 ° toward flexion/supination and then
moves back again to the maximum ex-
tension/pronation angle. The movement
through the final 10 ° is repeated five
times at a slow speed.
At the end of the cycle, the carriage will
again move to the maximum flexion
value, at the same time as the maximum
supination value, and then starts an-
other cycle with five repetitions through
the last 10 ° of extension/pronation.
This sequence can be repeated as often
as needed. You can stop treatment
with the STOP key; after the pro-
grammed therapy duration, it will stop
automatically.
Default: disabled
in the upper,
73
n EROM repeat
flexion/supination
The special "EROM repeat flexion/supi-
nation" function allows a more efficient
exercise in the last 10 ° before the set
maximum flexion and supination values.
For this protocol, the CPM device starts
in the middle between the set maximum
extension/flexion and pronation/supi-
nation values. First of all the carriage
moves simultaneously to the maximum
extension angle and to the maximum
pronation angle. Then it moves simulta-
neously to the maximum flexion angle
and to the maximum supination angle.
When the flexion/supination value has
been reached, the carriage reverses 10 °
toward extension/pronation and then
moves back again to the maximum
flexion/supination angle. The movement
through the final 10 ° is repeated five
times at a slow speed.
At the end of the cycle, the carriage will
again move to the maximum extension
value, at the same time as the maximum
pronation value, and then starts another
cycle with five repetitions through the
last 10 ° of flexion/supination.
This sequence can be repeated as often
as needed. You can stop treatment
with the STOP key; after the pro-
grammed therapy duration, it will stop
automatically.
Default: disabled
LEVEL 7:
n Transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Select the
function and press the START key. The
carriage moves to the transport posi-
tion. In the top left-hand corner of the
display you will see
to chapter 6 Transport).
n New patient
With this function, the CPM device will
move to the home position.
• Select the function and press the
START key:
– The carriage moves to the home
position.