Guidance and Manufacturer's Declaration – Electromagnetic Immunity
ARTROMOT®-E2/-E2 compact is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to ensure that the
ARTROMOT®-E2/-E2 compact unit is used in such an environment.
Immunity test
IEC 60601
test level
Home healthcare
Conducted RF to
6 V
IEC 61000-4-6
0,15 MHz-80 MHz
80% AM at 1 kHz
Radiated RF to
10 V/m
IEC 61000-4-3
80 MHz-2,7 GHz
80% AM at 1 kHz
Proximity fields
27 V/m
from RF wireless
380-390 MHz
communications
PM 18 Hz
equipment to
28 V/m
IEC 61000-4-3
430-470 MHz
FM +/- 5 kHz
1 kHz sine
9 V/m
704-787 MHz
PM 217 Hz
28 V/m
800-960 MHz
PM 18 Hz
28 V/m
1,7-1,99 GHz
PM 217 Hz
28 V/m
2,4 GHz-2,57 GHz
at PM 217 Hz
9 V/m
5,1 GHz-5,8 GHz
at PM 217 Hz
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the ARTROMOT®-E2/-E2 compact exceeds the ap-
plicable ARTROMOT®-E2/-E2 compact unit should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reori-
enting or relocating the ARTROMOT®-E2/-E2 compact unit.
Compliance
Electromagnetic
level
environment – guidance
Home healthcare
6 V
0,15 MHz-80 MHz
80% AM at 1 kHz
10 V/m
80 MHz-2,7 GHz
80% AM at 1 kHz
27 V/m
380-390 MHz
PM 18 Hz
28 V/m
430-470 MHz
FM +/- 5 kHz
1 kHz sine
9 V/m
704-787 MHz
PM 217 Hz
28 V/m
800-960 MHz
PM 18 Hz
28 V/m
1,7-1,99 GHz
PM 217 Hz
28 V/m
2,4 GHz-2,57 GHz
at PM 217 Hz
9 V/m
5,1 GHz-5,8 GHz
at PM 217 Hz
269