PCB Jumper Positions
Only the analog version of the Main board has a jumper on it. This section
provides the details for this jumper.
Main PCB Jumper
The IDNet version of the IND560 main board has no jumpers. There is one jumper
on the IND560 analog scale main board (W1). This jumper selects either 2 mV/V
or 3 mV/V operation of the analog circuitry. Refer to Figure 2-34 for the jumper
position.
When removing the W1 jumper for 3mV/V operation, reposition it on just one
of the pins as shown in Figure 2-34. If this jumper is not positioned properly,
the plastic cover over the Main board will not fit properly. If this occurs, do not
force the plastic cover down. Remove the cover and position the jumper as
shown.
The 2 mV/3 mV jumper settings are described in Table 2-4.
Jumper
W1
W1
Figure 2-34: Millivolt Jumper Selection
Table 2-4: Millivolt Jumper Description
On
Off
This jumper sets the connected scale's load cell setting to
X
2 mV/V.
No jumper sets the connected scale's load cell setting to
X
3 mV/V
IND560 Installation Manual
2mV/V
Description
3mV/V
2-21