10 General Setup (PATH: config/Guided Setup)
For Conductivity, the following additional options may be selected:
– USP (% safety margin below U.S. Pharmacopoeia limits)
– EP PW (% safety margin below European Pharmacopeia limits for Purified Water)
– EPWFI (%safety margin below European Pharmacopeia limits forWater for Injection)
– JP Cond (% safety margin below Japanese Pharmacopoeia limits)
10
From the main screen, press the Configure icon (C)
and select Guided Setup. From this screen, you will be
able to select and configure the first measurement from
the 5000TOCi that will be displayed by the M800, and
configure and activate an analog output and/or set point
to be associated with this measurement.
Analog outputs
By selecting Yes the linear 4–20 mA analog output Aout1
will be set up for measurement 1 when e is pressed.
Selecting No means that no analog output is set up.
Aout1 min, Aout1 max are the minimum and maximum
measurement values for the 4 and 20 mA values res-
pectively. Additional options may be configured from the
analog output configuration menu.
Set Points
On page 2 of the guided setup menu, a set point can
be configured for that measurement. Once the Set Point
option is set to Yes, then select the set point type you wish
to activate. If Off is selected, the set point will be disabled.
Once a set point type is selected, the necessary parame-
ters for that set point can also be configured.
The following set point types can be selected:
– High (High value has to be set)
– Low (Low value has to be set)
– Between (High and Low value has to be set)
– Outside (High and Low value has to be set)
After setting the set point value(s) a Relay can be selected
from the relays available on the transmitter, and configu-
red for that set point. The Relay delay is set to 10 seconds
and the Hysteresis is set to 5%. These parameters can
be configured from the Relay menu.