Data transfer
Wireless technology
Range
Frequency range
Modulation
Data rate (over the air)
Maximum output power (EIRP)
10.3 Directives and manufacturer's declaration
Tables according to IEC 60601-1-2 for the product 'BionicLink 60X3'.
Table 1 – Electromagnetic emissions for all devices and systems
Directives and manufacturer's declaration – electromagnetic interference
The product 'BionicLink 60X3' is designed for operation in an electromagnetic environment as
specified below. The customer or user of the product 'BionicLink 60X3' must ensure that the
device is operated in such an environment.
Interference measurements
HF emissions according to
CISPR 11
HF emissions according to
CISPR 11
Harmonics according to
IEC 61000-3-2
Voltage fluctuations/flicker
according to IEC 61000-3-3
Table 2 – Electromagnetic interference immunity for all devices and systems
Directives and manufacturer's declaration – electromagnetic interference immunity
The product 'BionicLink 60X3' is designed for operation in an electromagnetic environment as
specified below. The customer or user of the product 'BionicLink 60X3' must ensure that the
device is operated in such an environment.
Interference
immunity test
Electrostatic dis
charge (ESD) accord
ing to IEC 61000-4-2
Compliance
Group 1
Class B
N/A
N/A
IEC 60601 test level
± 6 kV discharge on
contact
± 8 kV discharge
through air
Bluetooth 2.0
approx. 10 m / 32.8 ft
2402 MHz to 2480 MHz
GFSK
704 kbps
-1.78 dBm (0.66 mW)
Electromagnetic environ
ment directive
The product 'BionicLink 60X3'
uses HF energy exclusively for
its internal functioning. There
fore its HF emissions are very
low and interference with
neighbouring electronic
devices is unlikely.
The product 'BionicLink 60X3'
is intended for use in other
than residential facilities and in
facilities connected directly to
a public low-voltage network
that also supplies residential
buildings.
Compliance level
± 6 kV discharge on
contact
± 8 kV discharge
through air
Electromagnetic
environment direct
ive
Floors should be
wood, concrete or
ceramic tile. If the
floor covering is made
of synthetic material,
the relative humidity
must be at least 30 %.
Ottobock | 23