AS 3000 IO-Link
Standard Variable "Standard Command" index=2 id=V_SystemCommand
data type: 8-bit UInteger
allowed values: 128 = Device Reset, 129 = Application Reset, 130 = Restore Factory Settings,
160 = Reset Primary Min/Max, 161 = Reset Secondary Min/Max, 164 = Reset All
Min/Max
access rights: wo
modifies other variables
octet
bit offset
element bit
Standard Variable "Device Access Locks" index=12 id=V_DeviceAccessLocks
data type: 16-bit Record (subindex access not
supported)
access rights: rw
subindex
bit offset
1
0
2
1
3
2
4
3
Variable "BDC1 Thresholds" index=60 id=V_Bdc1Thresholds
data type: 32-bit Record (subindex access not supported)
access rights: rw
modifies other variables
excluded from data storage
bit
subindex
offset
1
16
16-bit Integer
2
0
16-bit Integer
SP = switch point; RP = switch-back point; FL = level/temperature window lower value; FH = level/temperature window upper value
octet
0
bit offset
31 - 24
subindex
1
element bit
15 - 8
Variable "BDC1 Behavior" index=61 id=V_Bdc1Behavior
data type: 32-bit Record (subindex access not supported)
access rights: rw
modifies other variables
excluded from data storage
Stand: 20.04.2020
0
7 - 0
7 - 0
data type
name
Parameter (write)
Boolean
Access Lock
Boolean
Data Storage Lock
Local
Boolean
Parameterization
Lock
Local User Interface
Boolean
Lock
allowed
default
data type
values
value
See 11.2
350
See 11.2
400
1
2
23 - 16
15 - 8
1
2
7 - 0
15 - 8
HYDAC ELECTRONIC GMBH
description
Datenhaltung (Subindex 2) ist aus
Kompatibilitätsgründen vorhanden und sollte
immer auf „False" stehen, um unerwartetes
Verhalten beim Data Storage zwischen Master
und Gerät zu vermeiden.
For compatibility reasons only – it is
recommended to keep subindex 2 „false" in
order to avoid any unwanted system
behavior if data storage is set via IO-Link
master.
mod. other var.
V_PrimaryValueS1
Setpoint SP1
V_PrimaryValueSp1
Setpoint SP2
3
7 - 0
2
7 - 0
31
name
description
Lower limit of
RP (FL) in %
Upper limit of
SP (FH) in %
Mat. Nr.: 670006