8-channel Liquid End
Volume range[μl]
1 - 50
20 - 300
40 - 1000
Table 6: Accuracy Table 8-channel Liquid End
* Final test values relative to the nominal volume printed on the Liquid End (= max. volume) or the specified partial volumes at the same temperature (20 °C/68 °F) of the Liquid End, envi-
ronment within the Liquid Handling Station and aqua dest., as per DIN EN ISO 8655. A = Accuracy, VC = Variation Coefficient

2.6 Commissioning

2.6.1 Installation

Explosion hazard
Never operate the device in potentially explosive atmospheres.
The surface must be level and have a minimum carrying capacity of 25 kg for the Liquid Handling Station and
35 kg for the Liquid Handling Station flow.
Verify that the device is installed horizontally level with a spirit balance.
The recommended minimum installation space is H 700 mm / 750 mm x W 800 mm x D 600 mm. These di-
mensions guarantee a distance of approx. 100 mm of all sides of the device to the wall, thereby allowing un-
obstructed air circulation and protecting the device from overheating. The dimensions also allow 160 mm of
free space on top, which is required to open the front door.
Select a location that meets the following operating conditions:
+ 15 °C to + 35°C
max. 95 % relative humidity at 30 °C
Moreover, provide for sufficient space for a monitor and a PC or for a laptop.
NOTICE! Keep temperature differences and changes in humidity levels to a minimum during setup
and transport of the machine. Condensation can lead to malfunctions and defects in the Liquid Han-
dling Station or the computer connected to it.
Prior to commissioning, the Liquid Handling Station and the computer must first adapt to the ambient tem-
perature, which may require several hours under certain conditions.
Liquid Handling Station 9974 96
Partial volume [μl]
50
25
5
300
150
30
1000
500
100
DANGER
Devices and components
A* 
 %
VC* 
1.5
0.6
2.0
1.0
8.0
4.0
1.2
0.4
1.6
0.6
5.0
2.5
1.6
0.3
1.2
0.5
4
2
Operating manual English
 %
67

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis