• If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
Prolystica® 2x Neutral Detergent is validated for cleaning efficacy.
4
Low Level Disinfection (optional)
1.
Clean and prepare the camera head and cable as recommended in this
user guide. Ensure the soaking cap is installed.
2.
Disinfect the device in a disinfecting solution that has one of the
following active Ingredients:
• ≥ 2.4% glutaraldehyde
25°C
• ≥ 3.4% glutaraldehyde
25°C
• ≥ 0.55% ortho-phthalaldehyde
12 min at 25°C.
3.
Prepare the disinfecting solution according to the manufacturer's
instructions.
4.
Per manufacturer's recommendations, immerse the device, filling
all lumens, in the disinfecting solution for the required time at the
appropriate temperature.
5.
Dry all parts with a lint-free towel immediately after rinsing.
CIDEX Activated® is validated for disinfection efficacy.
5
CIDEX Plus® is validated for disinfection efficacy.
6
CIDEX® OPA is validated for disinfection efficacy.
7
Drying
• For automated drying, use the drying cycle provided with the
washer/disinfector.
• For manual drying, use a lint-free cloth.
• Dry any lumens with compressed air.
Maintenance, Inspection, and Testing
• Inspect the device on a continual basis. If a problem is observed or
suspected, the device should be returned for repair.
• Inspect all components for cleanliness. If fluid or tissue buildup is
present, repeat the above cleaning and disinfection procedures.
• Inspect the camera cable for cuts and breaks. Return any damaged
camera to Stryker for service.
with a minimum soaking time of 45 min at
5
with a minimum soaking time of 20 min at
6
with a minimum soaking time of
7
EN-19