P838 Auto False-Echo Suppression Distance
Defines the range of Auto False-Echo Suppression (P837) to use for ignoring false
echoes. (Units are defined in P005.)
Values
Determine the actual distance from the transducer face to the material surface. Subtract
0.5 m from this distance, and enter the result.
Set Up:
1. Perform this function when the vessel is empty or nearly empty.
2. Determine actual distance from transducer face to material level.
3. Select P838 and key in [distance to material level minus 0.5 m].
4. Press ENTER
5. Select P837.
6. Press 2 and then press ENTER
automatically after a few seconds.
Display after Auto False Echo Suppression
false
echo
7ML19985HT01
Range
(depends on model)
Default
.
TVT curve
material
(learned)
level
Distance (meters)
SITRANS Probe LU – INSTRUCTION MANUAL
Maximum range:
0.000 to 6.000 m (20 ft) or
0.000 to 12 m (40 ft)
1.000 m (3.28 ft)
. P837 will revert to 1 (use Learned TVT)