Maintenance
Adjusting the arm system
Adjusting the spring arm
The spring arm must hold the mounted display
mount in any height position.
Adjusting the height stop
Vertical movement of the display mount is limited
by a fixed lower stop and an adjustable upper stop.
The adjustable upper stop (height stop) can be set
to any position between the positions (A) and (B).
The factory setting for the spring arm is position (B).
A
1 Press down the seal inside the joint.
2 Adjust the height stop for the spring arm using
the set screw (C).
Use an Allen key (size 5 mm) for this.
Turn the set screw counterclockwise to adjust
the stop upwards.
Pay attention to the distance to the ceiling.
Turn the set screw clockwise to adjust the stop
downwards.
92
B
C
Adjusting the spring force
The spring force of the spring arm must be adjusted
if the display mount does not remain at the set
height.
When adjusting the spring arm, lift it approx. 10
above the horizontal position. Only in this position
is it possible to turn the set screw.
If necessary, set the upper stop (height stop) to a
higher position.
A
1 Raise the spring arm approx. 10
horizontal position.
2 Press down the seal inside the joint.
3 Adjust the spring force for the spring arm using
the set screw (A).
Use an Allen key (size 5 mm) for this.
Turn set screw counterclockwise to increase
the spring force.
Turn set screw clockwise to decrease the spring
force.
– If the adjustment has been carried out correctly,
the display mount can be positioned at any
height and will remain there.