Guidance and manufacturer's declaration – electromagnetic immunity
The AC-81MD/82MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the AC-81MD/82MD should assure that it is used in such an
environment.
IEC 60601
Immunity test
test level
Electrostatic
±8 kV contact
discharge
(ESD)
IEC 61000-4-2
±15 kV air
Electrical fast
±2 kV for
transient/burst
power supply
lines
IEC 61000-4-4
±1 kV for
input/output
lines
Surge
±1 kV line(s)
to line(s)
IEC 61000-4-5
±2 kV line(s)
to earth
Compliance
Electromagnetic environment – guidance
level
±8 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, a relative humidity of at
least 30% is recommended.
±15 kV air
±2 kV for
Mains power quality should be that of a
power supply
typical commercial or hospital
lines
environment.
±1 kV for
input/output
lines
±1 kV
Mains power quality should be that of a
differential
typical commercial or hospital
mode
environment.
±2 kV
common
mode
7
GB