Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data - VOLTCRAFT VC-5081 Bedienungsanleitung

Analog-multimeter
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Error
The multimeter is not working.
Resistance measurement not possible.
No measurement using the V/mA
socket (7) is possible.
No measurement using the 10 A
socket (4) is possible.
There is no change of the measured
value.
The multimeter is showing incorrect
readings.
Repairs other than the ones previously described should be carried out by an authorized
professional. If you have any questions concerning the operation of the multimeter, please do not
hesitate to contact our technical support.

13. Technical data

Measuring category ...........................................CAT III 500V against ground potential
Max. measuring voltage ....................................500 V
Analog display ...................................................with mirror scale
Input resistance V range ....................................approx. 10 MΩ
Power supply .....................................................9 V block battery + 2x AA
Operating temperature ......................................0 °C to +40 °C
Storage temperature ..........................................-10 °C to +50 °C
Operational height .............................................max. 2000 m
Rel. humidity ......................................................<75%, non-condensing
Temperature for guaranteed accuracy ...............+20 °C to +26 °C
Weight ...............................................................approx. 460 g
Measurements (LxWxH) ....................................160 x 105 x 40 (mm)
Possible cause
Is the power indicator on?
Are batteries used up?
Check the state of the 9 V battery and the fuse.
Are batteries used up? Check the state of both mignon (AA) batteries
and the fuse.
Is the fuse defective?
Check the fuse F 0.5 A/500 V (replacing the fuse).
Is the fuse defective?
Check the fuse F 10 A/500 V (replacing the fuse).
Are the correct measuring sockets selected?
Is the correct measuring function selected (AC/DC)?
Was the display zeroed (zero adjustment/zero ohm adjustment for
resistance measurement) before measuring?
Is the "Vrms/Vp-p" button (9) pressed?
45

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis