1. Important Information – Read Before Use
Read the safety instructions carefully before using the Ambu® aScope™ 4 RhinoLaryngo Slim.
The instructions for use may be updated without further notice. Copies of the current version
are available upon request. Please be aware that these instructions do not explain or discuss
clinical procedures. They describe only the basic operation and precautions related to the
operations of aScope 4 RhinoLaryngo Slim.
Before initial use of the aScope 4 RhinoLaryngo Slim it is essential for operators to have
received sufficient training in clinical endoscopic techniques and to be familiar with the
intended use, warnings and cautions mentioned in these instructions.
In this Instruction for Use, the term endoscope refers to instructions for aScope 4 RhinoLaryngo
Slim, and system refers to aScope 4 RhinoLaryngo Slim and the compatible Ambu monitor. This
Instruction for Use applies for the endoscope and information relevant for the system.

1.1. Intended Use

The endoscope is a sterile, single-use, flexible endoscope intended for endoscopic procedures
and examination within the nasal lumens and upper airway anatomy. The endoscope is
intended to provide visualization via a monitor.
The endoscope is intended for use in a hospital environment. It is designed for use in adults.

1.2. Warnings and Cautions

Failure to observe these warning and cautions may result in patient injury or damage to
the equipment.
Ambu is not responsible for any damage to the system or patient resulting from incorrect use.
WARNINGS
1. Only to be used by physicians, trained in clinical endoscopic techniques and procedures.
2. The endoscope is a single-use product and must be handled in a manner consistent
with accepted medical practice for such devices in order to avoid contamination of the
endoscope prior to insertion.
3. Do not soak, rinse, or sterilize this device as these procedures may leave harmful
residues or cause malfunction of the device. Reuse of the endoscope can cause
contamination, leading to infections.
4. Do not use the endoscope if the sterilisation barrier or its packaging is damaged.
5. Do not use the endoscope if it is damaged in any way or if the preuse check fails
(see section 4.1).
6. The images must not be used as an independent diagnostic of any pathology.
Physicians must interpret and substantiate any finding by other means and in the light
of the patient's clinical characteristics.
7. Do not use active endoscopic accessories such as laser probes and electrosurgical
equipment in conjunction with the endoscope, as this may result in patient injury or
damage to endoscope.
8. The endoscope is not to be used when delivering highly flammable anaesthetic gases
to the patient. This could potentially cause patient injury.
9. Patients should be adequately monitored at all times. Failure to do so may harm the patient.
10. Always make sure that the bending section is in a straight position when inserting and
withdrawing the endoscope. Do not operate the bending lever and never use excessive
force, as this may result in injury to the patient and/or damage to the endoscope.
11. Do not use excessive force when advancing, operating or withdrawing the endoscope
as this may result in patient injury or damage to the endoscope.
12. The distal end of the endoscope may get warm due to heating from the light emission
part. Avoid long periods of contact between the tip of the device and the mucosal
membrane as sustained contact with the mucosal membrane may cause mucosal injury.
4
CAUTIONS
1. Have a suitable backup system readily available in case a malfunction should occur.
2. Be careful not to damage the insertion cord or distal tip. Do not allow other objects
or sharp devices such as needles to strike the endoscope.
3. US federal law restricts these devices for sale only by, or on the order of, a physician.

2. System Description

The endoscope can be connected to the compatible monitor. For information about the
compatible monitor, please refer to its Instruction for Use.
2.1. System Parts
Part numbers:
Endoscopes
510001000 aScope 4 RhinoLaryngo Slim
aScope 4 RhinoLaryngo Slim is not available in all countries. Please contact your
local sales office.
Product Name
Colour
aScope 4 RhinoLaryngo Slim
Purple
Compatible monitors
Part numbers:
405002000 Model no. JANUS2-W08-R10
(SW versions ≥ V2.02)
®
aView
Ambu
TM
(Reusable)
For aView model no., please check the backside label on aView.
aView is not available in all countries. Please contact your local sales office.
Outer Diameter [mm]
min 3.0; max 3.5
5