Primo Star / Primo Star iLED
Fig. 46
Reflected-light LED module
• Attach the holder (Fig. 46/2) again, tighten the screws (Fig. 46/1) only slightly, still allowing
displacement of the LED module within the guide.
• Shift the LED module frontward (operator side) against the stop and fix the screws (Fig. 45/1)
handtight. This step is important to re-establish the optimally adjusted optical configuration.
• Plug the connector (Fig. 46/3) again into the circuit board.
• Put on the lid and tighten the three fastening screws (Fig. 45/1) after shifting the lid in direction of
arrow to the front right corner (Fig. 45) to realign the correct optical axis.
• Attach the tube again (see also section 3.7.1).
11/2013
Care and Troubleshooting
415500-7344-008
Carl Zeiss
55