SINGLE PATIENT CANISTER
SANITISATION
Before and after each use, sanitise the collection canister and the connecting tubing choosing one of the methods
described below.
(method A): rinse each part (6, 8, 9, 10, 10a, 11, 11a, 14) with hot potable water (approximately 40°C) with gen-
tle dish washing soap (non-abrasive) or in the dishwasher set to a hot cycle
(method B): immerse each part (6, 8, 9, 10, 10a, 11, 11a, 14) in a solution of 50% water and 50% white vinegar.
Lastly, rinse abundantly with hot potable water (approximately 40°C).
After having sanitised the accessories, shake them vigorously and place them on a paper towel. Alternatively, dry
them with a jet of hot air (for example, a hair dryer).
Reassemble the canister and the connecting tubing following the directions provided in the "CONNECTION DIAGRAM".
At the end of each use store the medical device complete with accessories in a dry place away from dust.
• The manual vacuum flow control (15) is a sterile disposable product and must be replaced after each use.
• ATTENTION: The hydrophobic/antiviral and antibacterial filter (7) supplied with Flaem aspirators must strictly be
replaced for each new patient, or if said filter is saturated.
Should the aspirator be used on the same patient, the filter approved for Single-patient use must be replaced every
TWO MONTHS at most or if the filter is saturated.
The filter cannot be sanitised, disinfected, or sterilised.
It is in any case good practice to sanitise and disinfect the device and the accessories, with the exception of the Filter,
always after each use, whether it is for the same patient or for a new patient.
STERILISABLE CANISTER
ACCESSORY SPECIFICATIONS
- hydrophobic antibacterial/antiviral filter (single-use)
- PC collection canister w/ compl. lid
- Ø 13 x 7.5 mm L 370 mm silicone tubing
- Ø 13 x 7.5 mm L 1300 mm silicone tubing
- Ø 13 x 7.5 mm L 80 mm silicone tubing
- manual vacuum flow control (single use)
MICROBIAL CONTAMINATION
In the presence of pathologies with the risk of infection and microbial contamination, we recommend personal use of the
accessories, collection canister and connecting tubing (consult your doctor).
Interactions:
The materials used for contact with the secretions are highly stable and chemical resistant thermoplastic polymers (PP,
PC, SI). We cannot, however, exclude interactions. Therefore, it is suggested: a) to always avoid prolonged contact of liquid
with the canister or tubing and sanitise immediately after use. b) Should anomalous situations occur, i.e. softening or
cracking of the accessories, quickly terminate the procedure and substitute the used materials. Contact the authorised
service centre and specify how the product was used.
The materials used in the device are biocompatible in accordance with the provisions of Directive 93/42 EC and
subsequent amendments. However, the possibility of occurrence of allergic reactions cannot be entirely exclud-
ed.
SINGLE PATIENT CANISTER
ACCESSORY SPECIFICATIONS
- hydrophobic antibacterial/antiviral filter (single use)
- PC collection canister w/ compl. lid
- Ø 13 x 7.5 mm L 250 mm silicone tubing
- Ø 10.5 x 9 mm PVC tubing w/fitt L 2000 mm
- Ø 13 x 7.5 mm L 40 mm silicone tubing
- manual vacuum flow control (single use)