CONFIGURATION

The X-M
AC radiographic system is factory confi gured in "standard mode".
®
IND
On the control panel the LED relevant to the following exposure parameters will light up:
No. of the selected tubehead
LED 1
Supplied cone
LED 8" = SHORT CONE
LED 12" = LONG CONE
Tubehead type
LED AC = ALTERNATING CURRENT
Radiographic voltage
LED 70 kV
Radiographic current
LED 8 mA
Type of patient
LED ADULT
Radiographic technique
CONVENTIONAL LED D
The following exposure times (sec) have been stored:
0.080 - 0.100 -0.125 - 0.160 - 0.200 - 0.250 - 0.320 - 0.400 - 0.500 - 0.630 - 0.800 - 1.00 - 1.250 - 1.600 - 2.000
- 2.500 - 3.200
PLEASE NOTE
These times are in compliance with current CEI EN 60601-2-7 (1999) norms and with the ISO 497 series R'10
recommendations.
MAY NOT BE MODIFIED
Certain exposure values have been predefi ned which depend on the selection of the operating parameters:
• Cone (8"/12")
• Type of patient (ADULT/CHILD)
• Radiographic technique
• Intra-oral test.
PLEASE NOTE
If one so desires, it is possible to change this values.