Herunterladen Inhalt Inhalt Diese Seite drucken

Siemens SITRANS F-Serie Bedienungsanleitung Seite 13

Ultrasonic flowmeters
Vorschau ausblenden Andere Handbücher für SITRANS F-Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

jurisdictional safety codes when operating this equipment. When properly installed the
equipment meets the following FM–CSA ratings:
Transmitter
● Explosionproof for Class I, Division1, Groups B, C, D;
● Dust-ignitionproof for Class II, Division 1, Groups E, F and G
● Intrinsically safe connections for Class I and II, Division 1, Groups A, B, C, D, E, F and G;
● Nonincendive for Class I, Division 2, Groups A, B, C and D;
● Suitable for Class II, Division 2, Groups B, E, F and G outdoor (Type 4X), Class III (CSA
Sensors
● Intrinsically safe Class I and II, Division 1, Groups A, B, C, D, E, F and G;
● Nonincendive for Class I, Division 2, Groups A, B, C and D;
● Suitable for Class II, Division 2, Groups B, E, F and G outdoor (Type 4X), Class III (CSA
● Temperature code T6 at an ambient of 40°C
ATEX installation
Read, understand and follow all safety instruction on the electronic media provided. This
equipment is rated for use in explosive atmospheres as stated below and must be installed
according to the 1010-422 installation drawing provided on the media. Failure to install the
equipment in the prescribed manner will result in unsafe operation. Follow all regional safety
laws when operating this equipment. When properly installed the equipment meets the
following ATEX ratings as stated in EC-Type Examination Certificate KEMA03ATEX2133
Transmitter
● IP65 – Ingress protection against solid bodies, rating of dust-tight and against liquid,
Sensors
● IP65 – Ingress protection against solid bodies, rating of dust-tight and against liquid,
FUH1010 IP65 NEMA 7 Interface Detector Quick Start
Operating Instructions, 06/2014, A5E02428148-AA
only)
only)
II 2 (1) G Ex d [ia] IIB+H2 – Category 2 Transmitter located in Zone 1 explosive
atmosphere with intrinsically safe circuits of category Ex ia, which can be connected to
Category 1 Sensors for use in potentially explosive atmosphere containing gases
rating of water jets
II 1 G Ex ia IIC T5 – Category 1 Sensors located in Zone 0 explosive atmosphere with
intrinsically safe circuits of category Ex ia for use in potentially explosive atmosphere
containing gases
rating of water jets
Introduction
1.2 Safety Notes
11

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

7me3601-1

Inhaltsverzeichnis