Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharge
as a result of low use of the instrument, even when
switched off).
Always remove the rechargeable batteries if you do not
intend to use the instrument for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.

7. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too weak.
weak
Re-position the cuff and repeat the measure-
ment.*
«ERR 2» Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle tension.
Repeat the measurement, keeping your
arm still.
«ERR 3» No pressure
An adequate pressure cannot be gener-
in the cuff
ated in the cuff. A leak may have occurred.
Check that cuff and pump ball are correctly
connected and that the cuff is not fitted too
loosely. Replace the batteries if necessary.
Repeat the measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
pressure too
300 mmHg) OR the pulse is too high (over
high
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats per
low
minute). Repeat the measurement.*
BP A50
* Please consult your doctor, if this or any other problem occurs repeat-
edly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.

8. Safety, Care, Accuracy Test and Disposal

Safety and protection

• This instrument may be used only for the purpose described in
this booklet. The manufacturer cannot be held liable for damage
caused by incorrect application.
• This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section!
• Protect it from:
− water and moisture
− extreme temperatures
− impact and dropping
− contamination and dust
− direct sunlight
− heat and cold
• The cuffs are sensitive and must be handled with care.
• Only pump up the cuff once fitted.
• Do not use the instrument close to strong electromagnetic fields
such as mobile telephones or radio installations.
• Do not use the instrument if you think it is damaged or notice
anything unusual.
• Never open the instrument.
• If the instrument is not going to be used for a prolonged period
the batteries should be removed.
• Read the additional safety instructions in the individual sections
of this booklet.
Ensure that children do not use the instrument unsuper-
vised; some parts are small enough to be swallowed.

Instrument care

Clean the instrument only with a soft, dry cloth.

Accuracy test

We recommend this instrument is tested for accuracy every
2 years or after mechanical impact (e.g. being dropped). Please
contact Microlife-Service to arrange the test (see foreword).
5