Herunterladen Inhalt Inhalt Diese Seite drucken

Measurements With Accessories - Chauvin Arnoux C.A 5220 G Bedienungsanleitung

Multimeters
Vorschau ausblenden Andere Handbücher für C.A 5220 G:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 50

MEASUREMENTS WITH ACCESSORIES

Different measurement accessories enlarge the field of application or add new
functions to your multimeter.
(See «To order» for the list of accessories and their reference. Documentation
on request)
HIGH CURRENT MEASUREMENTS
Miniclamps and current clamps, series C, D and Y, simplify the measurement
of AC currents.
Measurements are made rapidly and in complete safety, without the need to
switch Off the circuit.
According to the clamp used and the ranges concerned, you may measure
100A, 500A, 1000A and up to 3000A AC.
There are two types of clamps, with current output and with voltage output.
- Example of current output :
Output signal : 1mA AC/1 A AC
Use the mA AC ranges
For a current of 100A AC, reading 100mA AC
- Example of voltage output :
Output signal : 1mV AC/1 A AC
Use the mV AC ranges (C.A 5220 G only)
For a current of 100 A AC, reading 100mV AC
Hall effect clamps, series E and PAC, have the
same advantages but measure both DC and AC
currents
- output signal : 1mV DC/1 A DC and 1mV AC/1 A AC
MEASUREMENT OF HIGH DC VOLTAGES
The HT probe enables the measurement of high voltages up to 30kV DC and
20 kV AC
- output signal : 1V DC/1kV DC or 1V AC/1kV AC
LIGHTING MEASUREMENTS
The 651 A cell transforms your multimeter into a luxmeter; measurements up to
12000 lux (with 1/10 screen)
- output signal : 1mV DC/10 lux
TEMPERATURE MEASUREMENTS
The ST2 probe transforms your multimeter into a thermometer; it can be used in
many applications with 8 specialised sensors (type K); measurements up to
1000°C according to the sensor chosen.
- output signal : 1mV DC/1°C
MEASUREMENTS OF ROTATIONAL SPEED
The TACHY 6 probe transforms your multimeter into a tachometer; measurements
up to 6000 and 60000 rpm
- output signal :
1mV DC/1 rpm
1mV DC/10 rpm
29