Herunterladen Inhalt Inhalt Diese Seite drucken

Adjustment; Reference Phase - HBM MC2A Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

16

4.5 Adjustment

It is possible to adjust the measuring amplifier MC2A either with an electrical
signal (calibration box) or directly with a well defined transducer output signal.
After the zero adjustment a defined signal is fed in with a HBM calibrator, or
the transducer is mechanically detuned by some specific amount.
The range fine adjustment facility (screw driver potentiometer P 22) is then
used to set the corresponding output voltage (" 80 mV/V or " 8 mV/V
resp.¢ " 5 V).

4.6 Reference phase

Correction of the reference phase may be necessary when the apparatus is
first used or when long cables (e.g. > 20 m) are being used (due to the capac-
itance).
Correction of the reference phase is carried out by adjusting potentiometer
P 25. Having balanced the zero point connect a transducer whose plunger
can be displaced far enough to give an output signal of 4 ... 4.5 V.
Next, set the output voltage at maximum with P 25.
A signal for the reference phase setting can be attained by taking measure-
ments between point 9 and Kl. 23/8.
A correctly set reference phase is shown in the circuit diagram, Chapter 5.
HBM
MC2A
A1974−2.4 en/de/fr

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis