MA503/1-S
Parameter description
You can change or check the parameters at any time. The values you
selected are stored non-volatilely. For designation, function and select-
able values refer to the table in chapter 5.
To enter into programming mode
Press key
To leave programming mode
No key actuation for at least 30 s, or press key
parameter list is reached.
To scroll parameter information
Press key
To change parameters
Press keys
To store modified parameters
Press key
5 Parameter description
For a detailed parameter list containing all parameters that can be set and
the possibility to note down specific configurations refer to chapter 10.
(in English, parameter LAn = "E")
Display
"choice"
rES
FAc
MA503/1-S · Date 05.09.2016 · Art. No. 80994 · Mod. status 151/16
for at least 4 s.
.
and
.
, then message "-Sto-" will be briefly displayed.
Designation / description
Resolution: Determines the resolution of the display.
Parameter "FrEE" allows the programming of a calculat-
ing factor.
Calculation factor (only possible if resolution = 'FrEE'): The
calculation factor to be programmed is used for realizing
e. g. an angle display. Basis is the maximum possible res-
olution of 1/100 mm. The calculation factor "FAc" results
from:
FAc = meas. range / total travel distance [1/100 mm]
Example: Angle measurement on a circular disk with a dis-
play range of 0 ... 180°; display in 1/10°; circumference
of the circular disk 942.48 mm; hence total travel distance
471.24 mm;
FAc = 1800 / 47124 = 0.03820