(POS. 1.1) Nominal diameter of sensor
The following pages give a description of the most important functions and how they can be changed
or enabled/disabled by the user
(POS. 1.8) Electrodes cleaning
Selectable options: OFF, minimum, average and maximum. It is not recommended to use this
function (let OFF) when the liquid has a conductivity less than 100µS/cm.
(POS. 1.9) Empty pipe threshold
This value represents the threshold at which the measurement value is set to zero in case of empty
pipe condition; the span function is 20-250. Since the sensibility of empty pipe detection could be
changed by the liquid conductivity, ground connections, type of lining, in case of needed it is possible
to adjust manually the preset threshold. It is recommended to check periodically its proper
functioning. Value increasing means sensibility decreasing.
(POS. 1.10) Autozero calibration
This function activates the zero calibration automatic system. To perform this operation, it is
necessary for the sensor to be full of liquid, and the liquid be perfectly still. Even minimal movement
of the liquid may affect the result, and consequently, the accuracy of the system. Once you are sure
the above conditions have been fulfilled (and the percentage flow rate value is stable) press long the
key
and automatically activates the zeroing procedure, after which it is necessary to check that
the percentage value showed goes to zero, repeat the operation otherwise. When the value is stable
at zero, then press the key Enter.
(POS. 2.1-2) Full scale n° 1-2
The full scale is used to indicate to the maximum meter's flow rate; a volume per time is required.
The full scale should be chosen carefully as it's parameters are used for several other parameters
(Pls1-2, and 3 Measure). There are four fields to fill in order to set this parameter, from left to right:
1) volume unit of measure, 2) type of unit, 3) time unit of measure and 4) numeric value. The
selection is made by positioning the cursor on the field to modify. To change the type of unit of
measure (metric, British or American, mass or volume) the cursor has to be positioned on the symbol
"/" (field N. 2). When the nominal diameter is set to zero it is possible to modify only the numeric
field, since the unit of measure stays at m/sec. The following tables show the units of measure
available and the conversion factor by comparison with 1dm
kind of combination of units of measure satisfying both the following conditions:
Numeric field value  99999
1/25 fs
 numeric field value  fs
max

FUNCTIONS DESCRIPTION

(description of the functions with access code< 3)
MENU 1-SENSOR
MENU 1 - SENSOR
MENU 2 - SCALES
.
max
Identification of the function (not visualized on display)
[ND=
XXXX]
Menu visualized on the transmitter (from 1 to 11)
3
and 1kg. The transmitter accepts any
30
Transmitter request
Synthetic description of the function
[El. cleaning=MIN]
[E.P. thr.=XXX]
[Autozero cal.]
dm³/S
[Fs1-2=
X.XXXX]