- Do not screw or unscrew tips when the handpieces are on.
- All accessories must be cleaned, disinfected and sterilized before use.
- For your safety and that of your patient, I
than those supplied or recommended by SATELEC.
- Check the integrity of the device and its accessories before and after each use in order to detect
any problem.
- If necessary, do not use the device and replace any faulty component.
The I
C
2 is prepared for use as follows:
MPLANT
ENTER
- Make sure that the device is connected correctly and properly insulated.
- Switch the power switch (Fig. 2-5) to position I (ON).
- Suspend the bottles or bags of physiological saline or sterile water from the brackets.
- Raise the irrigation pump covers.
- Insert the irrigation line cassettes horizontally into their housings.
- Close the covers.
- Insert the irrigation line perforating pins into the bottles or pouches of physiological saline and open
the cap on the pins.
- Make any necessary setting adjustments (irrigation flow rate, speed, torque, power, power mode,
program, etc.) (refer to chapter 7).
Note: This sequence must be completed under conditions of controlled asepsis for the safety of your
patients. An operating assistant may be needed.
Note: The data is saved automatically, in order to maintain the display parameters as previously used
upon switchover from the various modes.
I-S
function
URGE
- Check that there are no traces of moisture on the I-S
connections. If there are, remove by wiping and drying with the multifunction syringe fed with medical-
quality filtered air.
- Connect the micro-motor cord to the lefthand connector of the device (Fig. 1-7).
- Align the micro-motor and cord connector electrical contacts, then screw the motor onto the
connector.
- Connect a contra-angle to the I-S
- Fit an instrument (drill, bur, etc.) to the contra-angle.
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the contra-angle.
- Run the motor fitted with its contra-angle and its instrument without irrigation at moderate speed for
10 to 15 seconds in order to diffuse the lubricant and eliminate excess.
- Prime the irrigation circuit by pressing the control keypad flush/prime button (Fig. 3-1).
- When the irrigation solution reaches the end of the contra-angle, release the footswitch.
- Use I
C
2 in accordance with good dental practice.
MPLANT
ENTER
P
function
IEZOTOME
- Connect the P
IEZOTOME
- Screw the selected tip onto the handpiece using a tip wrench (according to the recommendations of
the clinical booklets).
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the P
- Prime the irrigation circuit by pressing the flush/prime button on the screen or the multifunction
footswitch.
- When the irrigation solution reaches the end of the tip, release the footswitch.
- Use the I
C
MPLANT
ENTER
MPLANT
micro-motor.
URGE
2 handpiece cord to the righthand connector of the device (Fig. 1-9).
2 in accordance with good dental practice.
C
2 must not be used with accessories other
ENTER
URGE
2 handpiece.
IEZOTOME
micro-motor or micro-motor cord