Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 84

6.6.4. Auto Reverse Function

There are 3 different auto reverse modes:
AUTO REVERSING:
If, during operation the load reaches the preset torque limit value, the motor handpiece will
automatically rotate in the reverse direction. When the load is removed, the motor handpiece
returns to normal forward rotation automatically.
Load within the Torque Limit
Value
Forward Rotation
Fig. 8
AUTO STOP:
If, during operation the load reaches the preset torque limit value, the motor handpiece will
automatically rotate in the reverse direction. When the load is removed, the motor handpiece
stops. The LCD panel shows "
If you want the file to rotate forward again, press the ON/OFF key twice.
Load within the Torque Limit
Value
Forward Rotation
Fig. 9
AUTO REVERSE OFF:
If, during operation the load reaches the preset torque limit value, the motor handpiece will
stop without reverse rotation. The LCD panel shows "
alternately.
If you want the file to rotate forward again, press the ON/OFF key twice.
When the motor handpiece starts and its load reaches approximately half of the preset torque
limit value, the alarm sounds
The sound changes when the load approaches the torque limit value (corresponding to
on the bar display).
FISDR-JAP / F1902124.X / 00 / 2011 – created 07/2011
Further Loading
beyond the preset
Torque Limit Value
Reverse Rotation
" and the rotation speed alternately.
Further Loading
beyond the preset
Torque Limit Value
Reverse Rotation
(corresponding to
If the load is kept applied,
the file will rotate in reverse.
When the load is
removed,···
Forward Rotation
If the load is kept applied, the file
will rotate in reverse.
When the load is removed,···
Stop
" and the rotation speed
on the bar display).
- 24 / 208 -

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis