6 Troubleshooting

In the unlikely event that a problem occurs on the equipment, this troubleshooting guide can be used to identify the probable cause and remedy.
Troubleshooting should only be attempted by competent personnel. All major repair, and calibration work should be undertaken by a
Parker domnick hunter trained, qualified and approved engineer.
Fault
Power connected but the status
indicators and display (analyser only) is
not illuminated.
No / Low gas outlet pressure
High Oxygen concentration.
Low inlet pressure
Excessive noise or vibration
High outlet pressure.
Moisture in the gas at the outlet of the
generator.
Reduced flow at the outlet of the
generator.
Probable Cause
The electrical power to the generator is not
connected.
The electrical supply fuse is blown.
The controller ribbon cable is not connected.
External leak.
Internal leaks.
The pressure of the compressed air supply is
low.
The generator requires a service.
Defective Oxygen cell.
Leak in system piping.
The pre-filtration within the system is
approaching the end of its operational life.
The pretreatment dryer is being overflowed or
is operating at a reduced system pressure.
An isolation valve is partially closed upstream
of the generator.
External leak.
Silencer loose or defective.
Solenoid valve wear or coil loose.
Outlet regulator is incorrectly set or is
defective.
Blocked exhaust.
The CMS is beyond its operation life.
The flow controller is incorrectly set.
The dust filter on the receiver outlet is
blocked.
Faulty or incorrectly set pressure regulators.
Remedy
Check that there is power to the generator supply
terminals on terminal block "TB1".
Check fuse "F1" on terminal block "TB1".
If the fuse has blown disconnect the electrical supply
to the generator and replace the fuse.
Open the access door and check that the 26-way
ribbon cable is connected between the controller and
JP22 on the control board.
Check the piping and connection points for
leaks.Repair as necessary.
Open the access door and check all connection
points for leaks.
Repair as necessary.
Refer to Low inlet pressure fault below.
Check the service schedule and perform the
required service.
Replace the oxygen cell.
Open the access door and check all connection
points for leaks.
Repair as necessary.
Check the service shedules for the filters and
perform the required service.
Check that compressed air delivered to the dryer
meets the requirements specified within the
documentation provided with the dryer.
Check the position of all isolation valves.
Check the piping and connection points for
leaks.Repair as necessary.
Check that the exhaust silencer is securely fitted in
place.
Check the exhaust valves and verify that the coils
are secure.
Contact Parker domnick hunter for advice.
Contact Parker domnick hunter for advice.
Contact Parker domnick hunter for advice.
Contact Parker domnick hunter for advice.
Contact Parker domnick hunter for advice.
Contact Parker domnick hunter for advice.
Contact Parker domnick hunter for advice.