10. Alarm Display

Display
(Lights up)
(Flashing)
LH71A / LH72
Trouble
Measuring unit not
The measuring unit is not connected.
connected
Turn off the power, connect the measuring unit, and then turn
on the power again. The display value is reset to zero.
Excess speed
The maximum response speed is exceeded at the
measuring unit side.
Perform resetting operation.
(The same condition may occur when the machine is
subjected to a major shock.)
Overflow
When the display has overflowed, an "F" is added to the
highest digit.
Use in a range where an "F" is not added.
Power failure
The power fails momentarily during measurement.
Perform resetting operation.
Error in stored data
The stored data has been changed by noise or other cause.
Redo the settings starting from the basic settings.
If this error is displayed frequently, the memory may be
damaged. Contact your vendor.
Error in reference point
This is displayed when a measuring unit without a reference
detection
point is connected or when the reference point signal wire in
a measuring unit with a reference point is broken. Connect a
measuring unit with a reference point. If this does not correct
the problem, contact your vendor.
Program error
execution of a canned cycle.
Press the
operation to clear the error, and then run the program again.
Program error
There is no canned cycle during program execution or the
φ (diameter) display is set.
Check that any canned cycle programs are registered and
that the φ (diameter) display setting is not made.
Causes / Remedy
: Error code (1 to 9, A to F)
/
occurs during program execution or
key to close the program, perform resetting
10-1