Seite 1
Bedienungsanleitung Instruction Manual NOTE technique CDI01 Dosiergerät mit integriertem induktivem Leitfähigkeitsregler Metering Unit with Integrated Conductivity Control Unit Appareil de dosage avec commande de conductivité inductive intégrée 417101089_CDI01.doc - 1 - Rev.: 4-01/10...
Montage und Anschluss..............................9 Übersicht: „Wandmontage“ ..................................... 9 Übersicht: „Anschluss der Saug- und Dosierleitung an die Reinigerdosierpumpe“ ..................10 Übersicht: „Elektrischer Anschluss“................................10 5.3.1 Anschluss der Messzelle CDI01 (Artikel-Nr. 287410) ..............................11 5.3.2 Tankanschluss .......................................... 11 Installationsvoraussetzungen..................................11 5.4.1 Wandmontage ........................................... 12 5.4.2...
Seite 3
Activation of the external enabling signal for detergent metering (optional with second PCB, article-no. 223552) ........56 10.3 Activation of the pre-metering function (optional with second PCB and metering enable external to „YES“) ..........56 Technical data ................................57 Start up ..................................58 Calibration check on CDI01 ............................59 Faults ..................................60 Wearing and replacement parts..........................61 Appendix...................................62 16.1 Plans of terminal connections CDI 01 ................................
Seite 4
Tables des matiéres Informations générales .............................67 Entretien et réparation....................................67 Endommagements dus au transport ................................67 L´étendue de notre garantie..................................67 Adresse de contact / Fabricant ..................................67 Instructions de sécurité .............................68 Avertissements ......................................68 Enumérations........................................ 68 Instructions de sécurité spéciales lors des travaux de maintenance et de réparation................... 68 Contenu de la livraison .............................69 Composants................................70 Montage et connexions.............................71...
Sicherheitshinweis Bei Wartungs- und Reparaturarbeiten an Teilen, die mit Reiniger in Berührung kommen, sind wegen der Verätzungsgefahr geeignete Schutzkleidung und Schutzbrille zu tragen. Vor allen Reparaturarbeiten muss die Spannungsversorgung unterbrochen werden. Hervorhebungen Die Hervorhebungen „VORSICHT“, „ACHTUNG“ und „HINWEIS“ haben folgende Bedeutung in diesem Handbuch: wird benutzt, wenn ungenaues Befolgen oder Nichtbefolgen von Bedienungsanweisungen, VORSICHT...
Übersicht: „Anschluss der Saug- und Dosierleitung an die Reinigerdosierpumpe“ Pos. Bezeichnung Abb. 5.2 Obere Abdeckung Pumpenkopf O-Ring Schlauchadapter Klemmring Überwurfmutter Schlauchschelle Dosierschlauch untere Abdeckung Saugschlauch Überwurfmutter O-Ring Übersicht: „Elektrischer Anschluss“ Abb. 5.3 Pos. Bezeichnung Spannungsversorgung je nach Geräteausführung (240 V AC, 15 VA / 120 V AC, 30 VA / 24 V AC, 15 VA) Eingang Reinigerleermeldung Eingang LF-Messzelle Potentialfreies externes Freigabesignal für Dosierfreigabe und Vordosierung...
5.3.1 Anschluss der Messzelle CDI01 (Artikel-Nr. 287410) Einbau der Messzelle: Abb. 5.4 In die Tankwand (Pos. 6) ein Loch (ø 10 mm) bohren und mit Hilfe eines Rekordlochers auf ø 23 mm ausstanzen. Die Überwurfmutter (Pos. 1) lösen und gemeinsam mit dem O-Ring (Pos. 2) von der Anschlussleitung (Pos.
5.4.1 Wandmontage Entsprechend der örtlichen Wandbeschaffenheiten müssen passende ACHTUNG Befestigungselemente verwendet werden. Die Wandmontage ist wie folgt durchzuführen: 1. Die vier Deckelschrauben (Kapitel 5.1, Abb. 5.1, Pos. 3) entfernen und den Deckel (Abb. 5.1, Pos. 2) abnehmen. 2. Die Bohrlöcher an der vorgesehenen Stelle anzeichnen (Montageplatte als Bohrschablone verwenden).
Funktionsbeschreibung Spannungsversorgung und Freigabesignale Die Hauptplatine des CDI 01 wird über einen internen Trafo mit Spannung versorgt. Abb. 6.1 Pos. Bezeichnung Dosierpumpe mit integriertem induktiven Leitfähigkeitsregler Dosierleitung Saugleitung Leermeldung Sauglanze Vorratsbehälter Netzzuleitung Art. Nr. Bezeichnung 123519 & 240/ 2 x 12VAC, 15VA 123524 123525 &...
Aufbau der Steuerelektronik Das CDI 01 besitzt eine Hauptplatine (mit Mikroprozessor-Chip) und eine zweizeilige, 16-stellige Anzeigeeinheit (LC-Display). Alle Grundfunktionen des Gerätes werden von dieser Platine gesteuert. Die Bedienstruktur ist in die folgenden Ebenen unterteilt: • Betriebsebene • Serviceebene (SERVICE FUNKTIONEN) •...
Betriebsebene Innerhalb der Betriebsebene kann das Gerät ausgeschaltet werden. Von der Betriebsebene aus gelangt man über den Zutrittscode in die Service-, Parametrier- und Konfigurationsebene. Tastenfunktionen Service-, Parametrier- Betriebsebene Taste(n) und Konfigurationsebene Aktivierung des Menüs Aufrufen eines Menüs. Zutrittscode Rückkehren zur Ausgangsebene Anzeigen des unkompensierten Aufrufen eines Menüs.
7.2.2 Anzeige bei ausgeschaltetem Gerät: Durch gleichzeitiges Drücken der Tasten ↓, ↑ und ↵ für 5 Sekunden kann die Regelfunktion ausgeschaltet werden. 150 mS/cm 85°C Der aktuelle Leitfähigkeits- und Temperaturwert wird weiterhin angezeigt. Das Wiedereinschalten erfolgt durch Drücken einer beliebigen Taste. 7.2.3 Optionale Anzeigen: (Diese Anzeigen erscheinen nur in Verbindung mit einer...
Serviceebene In der Serviceebene (SERVICE-FUNKTIONEN) können: • Ausgänge aktiviert werden (MANUELLE FUNKTIONEN) • der Zutrittscode aktiviert, deaktiviert oder geändert werden (ZUTRITTSCODE-FUNKTIONEN). Die in den nachfolgenden Abschnitten dieses Kapitels aufgeführten Programmierschritte HINWEIS und Display-Texte sind in der Übersicht „Programmierstruktur“ im Anhang zusammenfassend dargestellt.
Funktionen für den Zugangsschutz Über das Menü ZUTRITTSCODE-FUNKTIONEN kann der Zutrittscode aktiviert, deaktiviert und geändert werden. Zur Aktivierung des Zutrittscodes ist folgendermaßen vorzugehen: 1. Das Menü ZUTRITTSCODE-FUNKTIONEN aktivieren. 2. Das Menü AKTIVIERE ZUTRITTSCODE aktivieren. 3. ↵ drücken und im Menü CODE EINGEBEN den gewünschten 4-stelligen Code eingeben.
Abgleich Funktionen 9.2.1 Anpassung der Leitfähigkeitsanzeigen Die Leitfähigkeits-Messzelle bzw. der Messverstärker sind ab Werk auf die Einhaltung bestimmter Fehlergrenzen überprüft. Wird trotzdem eine Anpassung der Leitfähigkeitsanzeige an bestimmte Vorgaben (Laborwerte) gewünscht, so kann dies mit der Funktion „LEITFÄHIGKEIT ± XX % durchgeführt werden.
Ferner kann eine „Verzögerungszeit für die Vordosierung“ (= verzögerter Start der Vordosierung) eingestellt werden. Es empfiehlt sich eine Verzögerungszeit zu programmieren, um zu verhindern, dass die Reiniger-Zudosierung vor der Anfangsbefüllung des Tanks in Gang gesetzt wird. Die einzustellende Vordosierzeit hängt vom Fassungsvermögen des Waschtanks, der Förderleistung des Dosiergerätes und der gewünschten Konzentration ab.
Störmeldeeinstellungen 9.4.1 Alarm bei Reinigermangel Dieser Alarm wird ausgelöst, wenn der Vorratsbehälter für den Reiniger leer ist, vorausgesetzt, eine Sauglanze mit Leermelde-Einrichtung ist angeschlossen. Für die Alarmanzeige und die externe Alarmmeldung kann eine Verzögerungszeit (REINIGER LEER VERZÖG.: 0005 s) eingestellt werden. Die Dosierung wird für die Dauer des Reinigermangels unterbrochen.
9.4.4 Alarm bei Reiniger-Überdosierung Mit der Festlegung eines Grenzwertes für die Leitfähigkeit (REINIGER ÜBERDOS. KONZ.: XXX mS/cm) kann verhindert werden, dass zuviel Reiniger in den Waschtank dosiert wird. Ein Überschreiten des festgelegten Wertes löst den Überdosierungsalarm aus. Durch Festlegung einer Alarmverzögerungszeit (REINIGER ÜBERDOS./VERZÖG.: 0120 s) kann verhindert werden, dass der Alarm bereits bei einer kurzfristigen, unbedeutenden Überdosierung ausgelöst wird.
Der Grenzwert für die Temperatur kann von AUS - 99°C eingestellt werden (ab Werk ist AUS eingestellt). Der Temperatur-Grenzwert ist wie folgt einzustellen: 1. Das Menü „TEMPERATUR-GRENZE MINIMUM:“ aktivieren. 2. Mittels der Pfeiltasten den gewünschten Wert einstellen. 3. Die Einstellung mit NEXT bestätigen. Die Verzögerungszeit kann von 0 bis 9999 s in Schritten von 1 s eingestellt werden (ab Werk sind 0120 s eingestellt).
Konfigurationsebene In der Konfigurationsebene (KONFIGURATIONS FUNKTIONEN) können folgende Eingaben vorgenommen werden: • Auswahl der Sprache (BEDIENERSPRACHE: DEUTSCH/ ENGLISH etc.) • Aktivierung der Reiniger-Nachdosierfunktion über externes Freigabesignal (optional mit Zusatzplatine) (DOSIERFREIGABE EXTERN ?) • Aktivierung der Reiniger-Vordosierfunktion über externes Freigabesignal (optional mit Zusatzplatine) (VORDOSIERUNG EXTERN ?) Die in den nachfolgenden Abschnitten dieses Kapitels aufgeführten Programmierschritte HINWEIS...
Technische Daten Spannungsversorgung: (230 V, 50/60 Hz), (115 V, 50/60 Hz), Je nach Ausführung (24 V, 50/60 Hz) Max. Stromaufnahme: 65 mA Spannungsversorgung Hauptplatine: 2 x 12 V AC, 15 VA 50/60 Hz Schutzart: IP 65 Schutzklasse: Dosierleistung: ca. 15 l/h gegen 0,6 bar Saughöhe max.
Inbetriebnahme Die Inbetriebnahme ist wie folgt durchzuführen: 1. Die Kastenwaschmaschine einschalten. 2. Die Konfiguration des Dosiergerätes durchführen. 3. Die Befüllung des Waschtanks mit klarem Wasser starten. 4. Alle Parametereinstellungen vornehmen. 5. Die Sauglanzen an das Reinigergebinde anschließen. 6. Nach 2 bis 3 Reinigungszyklen die Konzentration durch Titration überprüfen. Bei Bedarf den erforderlichen Leitfähigkeitswert entsprechend anpassen.
Überprüfung der Leitfähigkeits-Kalibrierung Am CDI01 kann die Leitfähigkeitsanzeige mit dem Leitfähigkeitssimulator für LMIT08, Artikel-Nr. 289190 oder mit der Kalibrier Box für Multronic, Artikel-Nr. 255195 überprüft werden. • Zur Überprüfung der Leitfähigkeits-Kalibrierung muss die Messzelle sauber und trocken sein (frei von leitfähigen Belägen).
Fehlersuche Der akustische Alarm kann durch Drücken einer beliebigen Taste an der Frontseite ausgeschaltet werden. Die Alarmanzeige im Display bleibt jedoch solange erhalten, bis der Fehler behoben ist. HINWEIS Ein “SYSTEMFEHLER“ wird erst nach aus- und einschalten des Gerätes quittiert. Displayanzeigen Fehlerursache Behebung...
• The unit is only used in accordance with the contents of this I & O Manual • Only original spare parts are used for service and repair • Only approved Henkel-Ecolab products are used. In addition, the general warranty and service conditions of the company ECOLAB-Engineering GmbH are applicable.
Safety instructions For maintenance and repair work of parts, which are in contact with detergents, it is imperative to wear the correct protective clothing and protective glasses, due to the risk of injury. Before any repair work the unit must be disconnected from the power supply. Special importance notes The expressions “WARNING”, “CAUTION”...
5.3.1 Connection of the conductivity sensor CDI 01 (article -no. 287410) Installation of the conductivity sensor: fig. 5.4 Drill a 23 mm hole into the tank wall (pos. 6). Release the union nut (pos. 1) and withdraw it together with the O-Ring (pos.
5.4.1 Wall mounting Ensure the correct mounting fixtures are used for the type of wall construction CAUTION (i. e. brick, block board)! Perform the wall mounting as follows: 1. Remove the four cover screws (chapter 5.1, fig. 5.1, pos. 3) and take off the enclosure cover (fig.
Functional description Power supply and enabling signals A internal transformer supplies power to the main board of the CDI 01. Pos. Designation fig. 6.1 Metering pump with integrated inductive conductivity controller Metering pipe Suction pipe Chemical empty signal Suction lance Chemical reservoir (drum) Main supply cable art.
Construction of the control electronics The CDI 01 comprises a mother board with a microprocessor and a dual line, 16-digit LC-display. All basic functions are controlled by this board. The structure of the electronic control is spilt into to the following levels: •...
Operation level The unit can be turned off even during operation. From the operating level you can gain access to the service-, parameter- and configuration level. Button functions Service-, parameter- and Operation level Button configuration level Call up the menu To call up a menu.
7.2.2 Display when unit is switched off: Press button ↓, ↑ together with button ↵ for 5 Sec. to stop the dosing control function. 150 mS/cm 85°C The current conductivity and temperature is shown in the display. Press any button to start the unit again. 7.2.3 Optional indication: (These indications are only available in connection with an...
Service level Within the (SERVICE FUNCTIONS) level the following functions are possible: • to activate the outputs (MANUAL FUNCTIONS) • to activate, deactivate or change the access code (ACCESS CODE FUNCTIONS). The programming steps as well as the display texts are shown as a complete block on the NOTE general overview of the programming structure in the appendix.
Access code functions Within the menu ACCESS CODE FUNCTIONS it is possible to activate, deactivate and change the access code. Perform the activation of the access code as follows: 1. Activate the menu ACCESS CODE FUNCTIONS. 2. Activate the menu ACTIVATE ACCESS CODE. 3.
Balancing functions 9.2.1 Conductivity indication adjustment The conductivity sensor and the measurement amplifier are tested to ensure correct measurement is obtained. Despite this fact that the actual conductivity is needed, the function “CONDUCTIVITY ± XX %” allows the setting of conductivity indication according to fixed instructions (lab values).
Furthermore, it is possible to set “delay for pre-metering” (= pre-metering starts with delay). It is recommended to programme a time for the detergent metering to be delayed before the first fill up. The pre-metering time depends on the capacity of the washing tank, the transferring capacity of the metering unit and the desired concentration.
Adjustment of fault messages 9.4.1 Alarm for detergent supply low This alarm is activated, when the detergent storage vessel is empty, via the connection of a suction lance with empty signal. The delay time for the alarm message and the external alarm is adjustable (EMPTY DETERGENT DELAY: 0005 s).
9.4.4 Alarm for detergent over dosage The adjustment of the conductivity limit (DET OVERDOSAGE CONZ.: XXX mS/cm) protects detergent overdosing into the washing tank. If the adjusted value is exceeded a over dosage alarm is activated. An adjusted alarm delay time (DET OVERDOSAGE DELAY.: 0120 s) prevents, the alarm activating for a short period of negligible over dosage.
The limit for the temperature can be adjusted from OFF - 99°C (pre-set from works: OFF). Adjustment temperature limit: 1. Activate the menu “TEMPERATURE MINIMUM:” 2. Select the desired value with button. 3. Press NEXT to confirm. The time delay can be adjusted from 0 to 9999 s in steps of 1 s (adj.
Configuration level In the configuration level (CONFIGURATION FUNCTIONS) the following inputs are possible: • Choice of language (MESSAGE LANGUAGE: GERMAN/ENGLISH etc.) • Start of the detergent pre-metering function with external enable (optional with second PCB) (METERING ENABLE EXTERNAL ?) • Start of the detergent pre-metering function with external enable (optional with second PCB) (PREMETERING EXTERNAL?) The programming steps as well as the display texts are shown as a complete block on the NOTE...
Technical data Power supply: (230 V, 50/60 Hz), (115 V, 50/60 Hz), depending on the version (24 V, 50/60 Hz) 65 mA Max. current consumption: Power supply main PCB: 2 x 12 V AC, 15 VA 50/60 Hz Ingress of protection: IP 65 Protection class: Metering capacity:...
Start up Start up the unit as follows: 1. turn on the machine washer 2. configure the metering unit 3. start the filling of the washing tank with clear water. 4. select all parameter adjustments 5. connect the chemical suction probe to the detergent vessel 6.
Calibration check on CDI01 The conductivity display of the CDI01 can be checked with the conductivity simulator for the LMIT08, article-no. 289190, or with the calibration box for the Multronic, article-no. 255195. • To check the calibration, the measuring cell has to be clean and dry (free of conductive fouling).
Faults The audible alarm can be switched off by pressing any button at the front. The alarm message on the display remains until the fault is eliminated. NOTE A “SYSTEM ERROR“ is only confirmed after turning the device on and off. display indication fault causes removal...
Seules les pièces de rechange d´origine sont utilisées. • Seuls les produits homologués doivent être utilisés. Par ailleurs, seront valables les conditions générales de garantie et de prestations de la société ECOLAB-Enigineering GmbH. Adresse de contact / Fabricant ECOLAB Engineering GmbH Raiffeisenstraße 7 D-83313 Siegsdorf Téléphone : (+49) 86 62 / 61 0...
Instructions de sécurité A cause des risques de brûlures il est absolument nécessaire de porter des vêtements protecteurs et des lunettes de protection appropriés chaque fois qu´il faut effectuer travaux d´entretien ou de réparation sur des pièces qui entrent en contact avec des produits de nettoyage.
Composants fig. 4.1 Nomenclature Rep. Boîtier Partie supérieure de la pompe “pompe de dosage de détergent“ Platine principale Couvercle Vis (4x) Témoin acoustique Transformateur 417101089_CDI01.doc - 70 - Rev.: 4-01/10...
Vue générale "Connexion des conduites d´aspiration et de dosage sur la pompe doseuse de détergent" Rep. Nomenclature fig. 5.2 Recouvrement Supérieur Partie Supérieure De La Pompe Joint Torique Adaptateur Du Tuyau Collier De Serrage Ecru-Raccord Bride Du Tuyau Tuyau De Dosage Recouvrement Inferieur Tuyau d´Aspiration Ecrou-Raccord...
5.3.1 Connexion du capteur CDI 01 (matériel n° 2.874.10) Montage du capteur fig. 5.4 1. Percer un trou (ø 10 mm) dans le paroi du réservoir (Rep. 6) et poinçonner jusqu´à ø 23 mm à l´aide d´un poinçon Record. 2. Détacher l´écrou-raccord (Rep. 1) et retirer conjointement avec le joint torique (Rep.
5.4.1 Montage mural ATTENTION Il faut utiliser des éléments appropriés de fixation selon la constitution du mur. Le montage mural s´effectue comme suit: 1. Déposer les quatre vis du couvercle (chapitre 5.1, fig. 5.1. Rep. 3) et enlever le couvercle (fig. 5.1, Rep. 2). 2.
Description fonctionnelle Alimentation électrique et signaux de libération L´alimentation électrique de la platine principale du CDI 01 s´effectue par un transformateur interne. Rep. Nomenclature fig. 6.1 pompe doseuse avec commande de conductivité inductive intégrée conduite de dosage conduite d´aspiration signal de niveau bas lance d´aspiration réservoir de stockage conduite au réseau...
Structure de la commande électronique Le CDI 01 est équipé d´une platine principale (avec puce de microprocesseur) ainsi que d´un élément d´indication à 16 unités (afficheur-LC). Toutes les fonctions de base de cet appareil sont commandées par cette platine. La commande électronique est divisée en les niveaux suivants: •...
Niveau d´opération L´appareil peut être arrêté dans le niveau d´opération. En utilisant un code d´accès vous arrivez du niveau d´opération aux niveaux de service, de paramètres et de configuration. Fonctions des touches Niveau de service, de paramètres Niveau d´opération Touche(s) et de configuration Activer le menu code Appeler un menu.
7.2.2 Indication lorsque l´appareil est arrêté: En pressant simultanément les touches ↓, ↑ et ↵ pendant 5 ARRET secondes, la fonction de régulation peut être éteinte. 150 mS/cm 85°C La valeur actuelle de la conductivité et de la température reste indiquée. Presser n´importe quelle touche pour enclencher l´appareil à...
Niveau de service Dans le niveau de service (SERVICE FONCTIONS) vous pouvez: • activer les sorties (MANUEL FONCTIONS) • activer, désactiver ou modifier le code d´accès (CODE D´ACCES FONCTIONS). Les pas de programmation indiqués dans le chapitre suivant ainsi que les textes INDICATION d´afficheur sont résumés à...
Fonctions pour la protection d´accès A travers le menu CODE D´ACCES FONCTIONS le code d´accès peut être activé, désactivé ou modifié. Pour activer le code d´accès il faut procéder comme suit : 1. Activer le menu CODE D´ACCES FONCTIONS. 2. Activer le menu ACTIVER CODE D´ACCES. 3.
Niveau de paramètres Réglages Dans le niveau de paramètres (PARAMETRES FONCTIONS) les réglages suivants peuvent être effectués : 9.1.1 REGLAGES FONCTIONS • Adapter la conductivité (CONDUCTIVITE +/- 0,0 % = XXX mS/cm) • Adapter l´indication de la température (TEMPERATURE 85 °C (0)) 9.1.2 DOSIER-EINSTELLUNGEN •...
Fonctions d´alignement 9.2.1 Adaptation de l´affichage de la conductivité Au moment de la fabrication à l´usine, le capteur de conductivité ou amplificateur de mesure sont contrôlés afin de respecter certaines limites de malfonctionnement. Dans le cas où une adaptation de l´indication de la conductivité à certaines conditions (valeur de laboratoire) est souhaitée, la fonction "CONDUCTIVITE ±...
De plus, il est possible de régler la “durée de retard du pré dosage“ (= retard du début de pré dosage). Nous conseillons de programmer une durée de retard pour éviter que le processus du dosage du détergent est activé avant le premier remplissage du réservoir. La valeur de réglage de la durée du pré...
Réglages des signaux de malfonctionnement 9.4.1 Alarme en cas de manque de détergent Cette alarme est activée si le réservoir de stockage est vide, et, pourvu qu´une lance d´aspiration équipée d´une installation du signal "niveau bas" est raccordée. Pour l´indication d´alarme et l´indication externe d´alarme il est possible de procéder au réglage d´une durée de retard (N.
9.4.4 Alarme en cas de surdosage de détergent La détermination d´une limite de conductivité (SURDOSAGE MAX. CONC. : XXX mS/cm) évite qu´une quantité trop grande de détergent est dosée dans le réservoir. En cas d´un dépassement de la valeur fixée, une alarme de surdosage est activée. La détermination d´une durée de retard d´alarme (SURDOSAGE MAX.
Le réglage de la valeur limite de température peut être effectué entre OFF - 99°C (au moment de la fabrication à l´usine la valeur est mis à OFF). Le réglage de la valeur limite de température s´effectue comme suit : 1.
Niveau de configuration Dans le niveau de configuration (CONFIGURATION FONCTIONS) les saisies suivantes peuvent être effectuées. • Sélection de la langue (LANGUES: ALLEMAND/ANGLAIS etc.) • Activer la fonction de dosage ultérieur du détergent à travers un signal externe de libération (optionnel avec platine additionnelle) (CDE EXTERNE DE DOSAGE?) •...
Spécifications techniques Alimentation de la tension selon modèle: (230 V, 50/60 Hz), (115 V, 50/60 Hz), (24 V, 50/60 Hz) 65 mA Consommation de courant max.: Alimentation de tension platine principale: 2 x 12 V AC, 15 VA 50/60 Hz Genre de protection: IP 65 Classe de protection:...
La mise en marche Pour la mise en marche de l´appareil il faut procéder comme suit : 1. Enclencher la laveuse à caisson. 2. Effectuer la configuration de l´appareil de dosage. 3. Activer le remplissage du réservoir avec de l´eau claire. 4.
Vérification de l’étalonnage de conductivité La vérification de l’étalonnage de conductivité du CDI01 peut être effectuée avec le simulateur de conductivité pour LMIT08, n° d’article 289190, ou avec la boîte d’étalonnage pour Multronic, n° d’article 255195. • Pour la vérification de l’étalonnage de conductivité, le capteur doit être propre et sec (exempt de tout dépôt conducteur).
Recherche de défauts En pressant n´importe quelle touche sur le côté frontal de l´appareil, le témoin acoustique peut être désactivé. L´indication d´alarme sur l´afficheur reste cependant active jusqu´à ce que remède a été porté au défaut. Un “ERREUR SYSTEME“ n’est confirmé qu’après avoir arrêté et mis en fonction INDICATION l’appareil.
Annexe 16.1 Plans d´affectation des bornes CDI 01 Plan de raccordement CDI 01 version de base 16.1.1 fig. 16.1 2.876.70 Rep. Nomenclature Platine de base Tension d´alimentation pour appareils de dosage du système Transformateur d´alimentation selon construction avec matériel n° 123519: 240/ 2x12 VAC, 15VA 123525: 120/ 2x12 VAC, 30VA 123527: 24/ 2x12 VAC, 15VA...
16.1.2 Plan de raccordement CDI 01 avec signal externe de libération "sans potentiel" fig. 16.2 2.876.70 Rep. Nomenclature Platine de base Tension d´alimentation pour appareils de dosage du système Transformateur d´alimentation selon construction avec matériel n° 123524: 240/ 2x12 VAC, 15VA 123526: 120/ 2x12 VAC, 30VA 123528: 24/ 2x12 VAC, 15VA Pompe doseuse de détergent...
16.1.3 Plan de raccordement CDI 01 avec signal externe de libération "12 – 230 VAC/DC " fig. 16.3 2.876.70 Rep. Nomenclature Platine de base Tension d´alimentation pour les appareils de dosage de l´unité Transformateur d´alimentation selon version avec matériel n° 123524: 240/ 2x12 VAC, 15VA 123526: 120/ 2x12 VAC, 30VA 123528: 24/ 2x12 VAC, 15VA...