JAGXTREME Division 2 and Zone 2/22 Installation Guide
Replacement Parts
Power Supply Battery
3-2
(7/03)
If a failure occurs in a JAGXTREME terminal that is installed in a Division 2 or Zone 2/22
area, note that only certain minimum revisions of some components may be used as
replacement parts. The following parts must be at least the letter revision indicated (or
newer), to be installed on a Division 2 or Category 3 approved JAGXTREME terminal. A
newer revision of a part number listed with a 'D" revision would begin with the letter 'E',
'F', 'G', etc.
Part Description
Controller PCB
Load Cell PCB (5V) (Single Channel)
Load Cell PCB (5V) (Dual Channel)
®
PROFIBUS
PCB
Analog Output PCB
AC Power Supply PCB
IF THE KEYBOARD, DISPLAY LENS OR ENCLOSURE IS
DAMAGED ON A DIVISION 2 APPROVED OR CATEGORY 3
MARKED JAGXTREME PANEL-MOUNT OR HARSH
TERMINAL THAT IS USED IN A DIVISION 2 OR ZONE 2/22
AREA, THE DEFECTIVE COMPONENT MUST BE REPAIRED
IMMEDIATELY. REMOVE AC POWER IMMEDIATELY AND
DO NOT REAPPLY AC POWER UNTIL THE DISPLAY LENS,
KEYBOARD OR ENCLOSURE HAS BEEN REPAIRED OR
REPLACED BY QUALIFIED SERVICE PERSONNEL. FAILURE
TO DO SO COULD RESULT IN BODILY HARM AND/OR
PROPERTY DAMAGE.
Note that an Alkaline battery (METTLER TOLEDO part number 14548600A) is used as
part of the power supply assembly. This battery has a nominal voltage of 4.5 volts DC
and a nominal capacity of 600 mAh (milliamp hours).
Part Number
C15686800A
E15360200A
F15360100A
B14688900A
D14095000A
C14200200A
WARNING