WHAT'S WRONG IF ...
... only a portion of the display lights up?
... the display is locked on one value or nonsense symbols are being displayed? – operating voltage switch incorrectly set
... the middle display sections blinks?
... the middle display sections will not stop blinking?
(longer than approx. 30 sec.)
... the green pilot lamp for the stability detector does not go out?
... only horizontal lines light up in the display during calibration procedure?
... CAL 100 blinks?
... CAL 0 blinks?
... CAL Err appears in the display?
... no CAL appears in the display?
THEN ...
– a momentary disturbance has occured (switch balance off and on again or
disconnect power cable and plug back in)
– DeltaDisplay is on.
– line voltage has dropped.
– the balance is in the configuration cycle (wait until 0.0000 appears).
– the weighing location or the load is too unstable (close sliding windows,
set longer integration time and/or switch off stability detector).
– the stability detector has been set too sensitively (see instructions under
"Integration time and stability detector")
– balance location is unfavorable (drafts, vibrations).
– the weighing location is too unstable (draft; close windows).
– the calibration lever has not been moved towards the rear, or an external
calibration weight has not been placed on the pan.
– the calibration lever has not been moved back to original position, or an
external calibration weight is still on pan.
– an object was on the weighing pan before beginning calibration procedure,
or an incorrect external calibration weight was used (return to weighing
mode by pressing and holding control bar).
– a momentary disturbance has occurred (recalibrate balance).